4 - 52
4.10 Treatment after Daily Use
1. Clean the processing unit.
Follow Steps 1 - 4 in “8.1 Wheel Dressing” to water the processing unit and wash out the
processing wastes with a brush.
CAUTION
• Be careful not to let water go into the inside of the instrument.
It may cause malfunction of the instrument.
NOTE
• Clean the processing unit after each use of the day. Be sure to clean it on the day that
the instrument is used.
If it is left for several days after use, the processing waste becomes settled and hard to
remove.
2. Turn the power OFF.
3. Perform after-use daily checks. See “4.11.2 Daily check after use” (p. 4-54).
4. Close the cover of the tracing unit.
5. Store accessories in a customary place.
Wipe the dirt on the accessories off and store them in a customary place to avoid possible loss
or breakage.