Quick guide Measurement and Calibration EC-meter EC-95 

Measurement: 
 

1) 

Insert the EC-sensor in the liquid. 

 

2) 

Press the 

MODE

 key to switch-on the instrument and the date 01/01/2020 of the last calibration/reset will be 

shown in the display followed by the actual EC-value of the liquid that is measured, wait until the value is 
stable and read of the value. 

  

 

 

 
EC calibration: 

 

        
When the sensor is dry and clean (not in a liquid), the instrument must show 0 (or close), when this is not the case, the user should first 
carry out an 0-calibration, check the normal user-manual for this procedure. 

 

 

1) 

Insert the EC-sensor in the Nieuwkoop EC-4 calibration liquid. 

2) 

Press the 

SET UP

 key for 5 seconds and 

SENS CAL

 will appear in the display, alternate to the 

 

       EC-value. 

 

 

 

Note: 

Wait until the value is stabilised!. 

 
      

   

 

 

 

 

3) 

Press the 

SET UP

 key and the standard calibration value 

4.00

 appears in the display. 

4) 

Press the 

SET UP

 key for approx. 5 seconds to confirm this value and 

UPDATE

 appears in the display. 

5) 

When the value is confirmed LAST CAL will appear in the display, followed by XX / XX / XX (this is the date 
of the last calibration), this date can be registered as follows: with the 

INCR

 key you can be change the 

figure and with SHIFT key  you can change the position of the cursor, then press the 

SET UP

 key for 5 

seconds to confirm the date and UPDATE will appear in the display. 

 

When the calibration liquid is not a standard Nieuwkoop 4.00 mS/cm value, the EC-95 EC-meter can be calibrated 
manually, read the normal user-manual for this procedure. 

 

A complete user-manual can be downloaded from the website www.meten.nl

 

 

 
 
 

Quick guide Measurement and Calibration EC-meter EC-95 

Measurement: 
 

3) 

Insert the EC-sensor in the liquid. 

 

4) 

Press the 

MODE

 key to switch-on the instrument and the date 01/01/2020 of the last calibration/reset will be 

shown in the display followed by the actual EC-value of the liquid that is measured, wait until the value is 
stable and read of the value. 

  

 

 

 
EC calibration: 

 

        
When the sensor is dry and clean (not in a liquid), the instrument must show 0 (or close), when this is not the case, the user should first 
carry out an 0-calibration, check the normal user-manual for this procedure. 

 

 

6) 

Insert the EC-sensor in the Nieuwkoop EC-4 calibration liquid. 

7) 

Press the 

SET UP

 key for 5 seconds and 

SENS CAL

 will appear in the display, alternate to the 

 

       EC-value. 

 

 

 

Note: 

Wait until the value is stabilized!. 

 
      

   

 

 

 

 

8) 

Press the 

SET UP

 key and the standard calibration value 

4.00

 appears in the display. 

9) 

Press the 

SET UP

 key for approx. 5 seconds to confirm this value and 

UPDATE

 appears in the display. 

10) 

When the value is confirmed LAST CAL will appear in the display, followed by XX / XX / XX (this is the date 
of the last calibration), this date can be registered as follows: with the 

INCR

 key you can be change the 

figure and with SHIFT key  you can change the position of the cursor, then press the 

SET UP

 key for 5 

seconds to confirm the date and UPDATE will appear in the display. 

 

When the calibration liquid is not a standard Nieuwkoop 4.00 mS/cm value, the EC-95 EC-meter can be calibrated 
manually, read the normal user-manual for this procedure. 

 

A complete user-manual can be downloaded from the website www.meten.nl

 