background image

                

TESTERS

 

EC2620 / EC2630 / EP2000 / PH2620 / PH2630 / PH2720

                

_____________________________________________________________________________________________

 

8

 

 

 

CALIBRATION PROCEDURE FOR CONDUCTIVITY (EC2620-Cond1) 

   

1.  Power ON the meter by pressing the        key. 

   2. Rinse the probe with distilled water. 
   3. Immerse the probe in the calibration solution (1413µS or 12.88 mS), wait for stability 

 

   4. Press 

CAL

 key. 

   5. The meter will start calibration procedure and will recognise automatically the  
       standard used. 
   6. When stable press          to confirm and complete the calibration. 
   7.  The standard value will flash for 3 times and then the meter will go into the 
        measurement mode. 
   8. If the 2

nd

 point calibration is required then rinse the probe with distilled water and 

       immerse in the second standard solution (1413µS or 12.88 mS), wait for stability 

 

   9. Repeat the points 4 to 7. 
  10. The calibration process is completed and the meter is ready to use. 
NOTE: Anytime press        key to abort and exit from calibration procedure. 

 

 

 

       Reading based on theoretical cell value C=1 
       Standard solution 
 

 

CALIBRATION PROCEDURE FOR TDS (EC2620-Cond1

)  

When the meter is set to read 

TDS

 then the calibration is done on TDS with 1 or 2 points.  

The calibration procedure for TDS is same as for Conductivity. 
 

 

CALIBRATION PROCEDURE FOR pH (PH2620-pH1)

 

 

    1.   Power ON the meter by pressing the        key. 
   2.   Rinse the electrode with distilled water. 
   3.   Immerse the electrode in the  1

st

 buffer solution pH7.0 and wait for stability 

  

   4.   Press 

CAL

 key. 

   5.   The meter will start calibration procedure and will recognize automatically the  
         standard used. 
   6.   When stable press           to confirm and complete the calibration. 
   7.   The standard value will flash for  3 times and then the meter will ask for next point  
         for calibration. If only 1 point calibration is required then press      to finish and exit. 
   8.   If the 2

nd

 point calibration  is required then rinse the electrode with distilled water 

         and immerse in pH4.0 or pH10.0, wait for stability 

 

   9.   Repeat the points 4 to 7. 
  10.   The calibration process is completed and the meter is ready to use. 

 
 

NOTE: Anytime press        key to abort and exit from calibration procedure. 
NOTE (2): When the first point calibration is confirmed (point 7) if the sensor is not  
removed from the buffer solution, the instrument may give wrong buffer error. 
 
 

Summary of Contents for EC2620

Page 1: ...USER MANUAL Testers EC2620 EC2630 EP2000 PH2620 PH2630 PH2720...

Page 2: ...PLY 6 INSTRUCTIONS FOR PH2620 pH1 EC2620 Cond1 7 POWER ON 7 POWER OFF 7 SETUP MENU for PH2620 pH1 EC2620 Cond1 7 MEASUREMENT 7 CALIBRATION PROCEDURE FOR CONDUCTIVITY EC2620 Cond1 8 CALIBRATION PROCEDU...

Page 3: ...TESTERS EC2620 EC2630 EP2000 PH2620 PH2630 PH2720 _____________________________________________________________________________________________ 3 PRODUCT SPECIFICATIONS...

Page 4: ...3 key for all functions SAFETY INSTRUCTIONS Read this instruction manual carefully before using your new tester The membrane of pH electrode is made of glass and can be danger in case it breaks To av...

Page 5: ...______________________________________________________________________ 5 PRODUCT DESCRIPTION KEYPAD PH2630 pH5 EC2630 Cond5 EP2000 PC5 PH2620 pH1 EC2620 Cond1 Keypad Functions for Tester PH2630 pH5 EC...

Page 6: ...CALIBRATION POINTS INDICATOR When a calibration is performed these icons indicate the points calibrated and First point for pH calibration is always 7 00 pH POWER SUPPLY This tester series works with...

Page 7: ...o select the parameter between COND or TDS to be used for measurements and press to confirm only for EC2620 Cond1 o Only if the TDS is selected then the display will show TDS FCT flashing press to cha...

Page 8: ...d solution CALIBRATION PROCEDURE FOR TDS EC2620 Cond1 When the meter is set to read TDS then the calibration is done on TDS with 1 or 2 points The calibration procedure for TDS is same as for Conducti...

Page 9: ...2630 Cond5 EP2000 PC5 1 With meter switched off press and hold CAL key and press key once 2 The meter will switch on with all the segments active release CAL key the meter will got into the Setup Menu...

Page 10: ...tandard value will flash for 3 times and then the meter will go into the measurement mode 8 If the 2nd point calibration is required then rinse the electrode with distilled water and immerse in the 2n...

Page 11: ...bration REPLACEMENT OF SENSOR Tester 5 series has replacement sensors which can be replaced in case it expires of damages 1 To replace the sensor unscrew the dial in anti clock wise 2 Pull out sensor...

Page 12: ...suring value is not stable during calibration Press key when icon appears During calibration the measuring value is not stable for 3min 1 Check whether there are bubbles in glass bulb 2 Replace with n...