-16-
[Leveling]
Leveling in one direction:
(1) Place the device on the object under measurement. Check the position of the bubble in the
auxiliary bubble tube, and perform 0-Call, so that the indicator will show zero.
(2) Turn the device around 180º, and check to make sure that the bubble in the auxiliary
bubble tube stays in the same position. Then, perform 1/2-Call to halve the indication
value. If the position of the bubble in the auxiliary bubble tube is different, adjust the object
under measurement, since it is likely that an error may be produced due to the inclination
in the roll direction (direction perpendicular to the measurement axis).
(3) Adjust the inclination of the object under measurement, so that the indication will become
zero.
(4) Turn the device around 180º once again, and check to see if the indication is zero or not. If
it is zero, it means that leveling has been accomplished.
If the indication is not zero, freshly redo Steps (1) through (4).
0-Call
1/2-Call
Adjustment