-19-
Signal output is basically controlled with CTS.
CTS is an instruction signal that tells the device from the outside whether it should deliver or not
deliver data.
When function selector switch is set in 1/2:
If the CTS terminal is at High Level, measurement data is output through the TD terminal.
If the CTS terminal is at Low Level or not connected, no measurement data is output.
If the CTS terminal is continuously at High Level, output takes place consecutively every time
the data is updated.
When function selector switch is set in OUT:
When the signal output switch (serving also as the 1/2-Call switch) is pressed, with the CTS
terminal at High Level, measurement data is output through the TD terminal. The signal
output switch is designed to deliver measurement data the moment it is released after being
depressed, so that two or more measurement data will not be sent out at a time in each
operation. Holding down the signal output switch continuously will not prompt measurement
data to be output.
(Symbol
denotes a space.)
Output in the unit of mm/m
Output in the unit of mm/m
Output in the unit of mm/m
Output in the unit of DEG (º)
Output in the unit of DEG (º)
Output in the unit of DEG (º)
Error output
Error output