X K 3 1 9 0
–
A26
()
14
scoreboard connect line(blue line)
;
Foot 5 of the plug on the back of the indicator is GND,connect with the scoreboard
connect line(black line).
(3)Parameter setting
①
Press"1"and "Clear"button in same time in weighing state.
Indicator display
【
bL
】
(
Power-saving mode chosen
)
【
*
】
Press
“
clear
”
enter into next optional function chosen.
②
Indicator display
【
Func
】
(
optional function chose
)
【
*
】
(
0~4
)
Now input
“
3
”,
after choose the scoreboard output extension function then
press"clear"button confirm and then enter to the next optional function chosen.
③
Indicator display
【
DISP
】
(
single window and three windows chosen
)
【
*
】
(
1
、
3
)
Now input single window
(
1
)
,three windows
(
3
)
after chosen and press
“
clear
”
button
confirm and indicator back to weighing state.
(4)Upper and lower limit alarming
<1>Brief introduction
A25 provide two point outputs of the alarm function (alarm mode can be set),
output method is transistor OC gate output (see below diagram), each of the
maximum drive current is 50mA.
<2>Wiring method
Air plug on the back of the
indicator is switch volume
signal output, which are
:
Foot 1 is switch volume 1output
;
Foot 4 is switch volume 2 output
Foot 5 is GND
;
<3>Parameter setting