Step 9.:
Log on to the network and start NIS elements. Please NEVER save data to the
local hard drives as they might be deleted!!
System Shut Down:
Step 1.:
ONLY proceed if you are the last user or if the instrument is not booked for an
hour or more.
Step 2.:
ALWAYS switch back to the 4X objective and distance the objective from the
stage. If the 60x oil immersion lens has been used NEVER forget to clean with lens cleaning
paper: 1. Fold the lens cleaning paper three times to form a strip, 2. Hold it on the two ends
on both sides, 3. Gently wipe with left-right motion. Never press with your finger directly.
4. Repeat previous step until you cannot see oil stain and repeat once after that one final
time.
Step 3.:
Stop NIS elements. If you proceed with the shutdown process without exiting NIS
Elements it will prompt you with error messages.
Step 4.:
Switch the epifluorescent lamp OFF.