VRA53801-R.3609.A
-
A48
・
-
LS-50
Rewrite IR pass filter compensation data
※
When the main PCB or parts of optical system are replaced, rewrite the compensation data according to the following
instrucition.
②
Select "Run" on the main menu, then choose "IR Leakage Intensity" and "Write" from the pulldown menus.
③
Input each data (Leak_g
・
Leak_s
・
Leak_k) that was recorded in
①
, then click "OK".
①
After replacing parts, perform "IR Leakage Test" (IR pass filter) from page A17 to A19. Record each data on inspection
results (Leak_g
・
Leak_s
・
Leak_k).