Chapter 1
Microscopy Procedures
22
Chapter 1-4
Micr
oscopy
Pr
oced
ures
Sensitive T
int
Plate
Mi
croscopy
4.2
Sensitive Tint Plate Microscopy Procedure
In this procedure, only step titles are shown for operations that are the same as those of bright-field microscopy. See “2.1
Bright-field Microscopy” for details.
1. Turn on the power.
2. Lower condenser slightly from
uppermost position.
3. Fully open field and aperture
diaphragms.
4. Bring the 10x objective into the optical
path.
5. Bring specimen into optical path.
6. Remove the analyzer from the optical
path.
7. Focus on specimen.
8. Adjust
diopter.
9. Adjust
interpupillary.
10. Focus and center condenser.
11. Bring the portion without sample into
the optical path.
12. Install polarizer unit.
13. Adjust orientation of analyzer and
polarizer.
14. Bring the desired objective into the
optical path.
15. Focus on specimen.
16. Circumscribe field diaphragm to field
of view.
17. View specimen.
18. Turn off the power.
CLAMP
T
O
RQ
UE
ND4
ND8
OUT
IN
ND4
ND8
OUT
IN
Preparation for microscopy
1
Turn on the power.
(The power LED on the front of the main body will light
up.)
2
Lower the condenser slightly from the uppermost position.
3
Fully open the field diaphragm and aperture
diaphragm.
4
Bring the 10x objective into the optical path.
2
5
8
9
4
10
3
6 12 13
14
1
7 15
11
16
17
18
10
6 12
3