45
Assembly
Assemble each part of the microscope by referring to the diagram on the next page.
WARNING
•
Before assembling the microscope, be sure to read the
WARNING and
CAUTION
at the beginning of this instruction manual and follow the instructions written therein.
•
To prevent electrical shocks and fire, turn off the power switch (flip it to the “ ” side)
when assembling the microscope.
CAUTION
•
Be careful not to pinch your fingers or hands during assembly.
•
Scratches or fingerprints on the lens surface will adversely affect the microscope
image. Be careful not to scratch or touch the lens surfaces. If lenses are contaminated
with fingerprint or such, clean them according to the procedure described in “VII.
Care and Maintenance.”
•
The microscope is a precision optical instrument. Handle it carefully and do not
subject it to a strong physical shock. (In particular, objectives may loose accuracy
when exposed to even a weak physical shock.)
Required tools
•
Hexagonal screwdriver
2 mm
×
2 (supplied with the microscope)
•
Hexagonal wrench
3 mm
×
1 (supplied with the microscope)
When not using, place these in the tool holder at the right side of the microscope base.
Installation location
Being a precision optical instrument, this product may get damaged or loose accuracy if it is used or
stored under unsuitable conditions. When selecting the installation location, note the following:
•
Avoid a brightly lit location, such as exposed to direct sunlight or directly under a room light.
The image quality deteriorates if there is excessive ambient light.
•
Choose a location that is free from considerable dust or dirt.
•
Choose a flat surface with little vibration.
•
Choose a sturdy desk or table that is able to bear the weight of the instrument.
•
Do not install the microscope in a hot or humid location.
•
Take enough space around the microscope referring to the layout diagrams on page 6.
•
The microscope may be moved by earthquakes. We recommend taking anti-earthquake
measures.
For details about anti-earthquake measures, see “15. Countermeasures for Earthquakes.”
•
For details about the operating environment and storage environment, see “VIII.
Specifications.”
Combination of the illuminator and the light source
This microscope system is approved by TUV and SEMI only in the combination of the illuminator
and the light source describe below. Please take note that if a light source other than the specified
ones are installed onto this microscope, this microscope system will not be treated as a TUV/SEMI
approved product.
•
Illuminator
Nikon LV-LH50PC precentered lamphouse 12V 50W
•
Lamphouse
Nikon LV-LH50PC precentered lamphouse 12V 50W
•
Lamp
Nikon LV-HL50W 12V 50W LONGLIFE halogen lamp or non-Nikon 12V
50W SHORTLIFE halogen lamp (model OSRAM HLX 64610, OSRAM HLX
64611, or PHILIPS 7027)