8

X) Remove cables and re-attach case and four screws. 
 
Y) test again boot process and with NikonScan 
 
Z) test again a week later and / or burn-in for 48 hours  
 

Appendix A:

  

(to remove the black mechanism assy from the metallic-colored case structural assy) 
 
This additional disassembly may be required to remove the top light source module to clean front 
surface mirrors, both on top and bottom where the lens and sensor module is. 
 
i) remove the four screws on top module, place in vial #5 
 
ii) remove tiny black screw on front corner of structural assy, place in vial #6. Note the indexing of 
the top plate with other screw heads 
 
iii) remove three screws on bottom that attach the black mechanism to base of the structural assy, 
place in vial #6  
 
The whole black mechanism will slide out. This will facilitate cleaning access to especially dirty 
mirrors and lens 
 
iv) reverse procedure for re-assembly. The light source assembly is in elongated sliding slots. 
Fasten in the mid-position. 
 
 

IV. Warranty and Limits of Liability 

 
This procedure is provided on a "best-efforts" basis and there is no warranty whatsoever herein, 
and is provided "AS-IS." It requires a high-level of technical competence and mechanical dexterity, 
such as an "A+ Certification" Computer Technician plus 3 years experience, or equivalent, and 
demonstrated ability to handle and understand complex, delicate optical mechanisms and 
instruments. 
 
(ie practice of something simpler, like reviving a bad CD-ROM, CD player or VCR that is eating 
tapes). 
 
A prerequisite is being able to comfortably use Scott Mueller's "Upgrading and Repairing PCs," the 
1500-page defacto "Bible" by Que, documentation available from manufacturers on the internet, 
and having successfully built a custom-made computer from scratch (separate components). 
 

This is the Entire Agreement between the User and the Service Provider 

 
The User assumes all risk in using this document. If this procedure is used, the User specifically 
agrees to "hold harmless" from any and all liability alleged the responsibility of the Author, which 

Summary of Contents for LS-2000

Page 1: ...s steady Position unit in vertical position with power switch on top 2 during the blinking phase the unit is booting up and goes through mechanism qualification steps see if you can hear any buzzing where stepper motors are moving the scanner mechanism Testing of Autofocus is first and X axis scan is second 3 look inside the slide module throat and see if you can see any narrow strip of light from...

Page 2: ...d so precisely that screws seem to stick on the tip without being magnetic May need a medical type hemostat forcep or tweezer to pick out dropped and re position tiny screws Available at an unusually complete Ace hardware store such as Seven Corners Hardware St Paul Minnesota 651 224 4859 7corners com with an upstairs nationally rated catalog mail order operation at 800 328 0457 2 DuPont Teflon Mu...

Page 3: ...flashlight 9 6v Ni Cd with ratcheting right angle head or equivalent Perhaps a Petzl Tikka forehead strapped LED flashlight will work too 4 Plastic pink antistatic foam pad 1 ft x 18 x 1 2 thick preferably grounded with 2 Meg ohm resistor 5 Cheseborough Ponds cotton Q tips or TexWipe foam swabs and aqueous lens cleaner isopropyl alcohol or Everclear liquor 190 proof ethanol Kimberly Clark Chemwipe...

Page 4: ...connectors put in vial 2 The circuit board should be loose but still attached E remove 5 cables on bottom side of circuit board and 2 cables on top side of circuit board May need to use fingernail thin jewelers screwdriver or forcep to separate the cable jack and connector Refrain from pulling on the wires with fingers or forcep just get it to barely separate then insert and twist a flat bladed sc...

Page 5: ...eft side especially for any bloated or leaking round cylindrical brown capacitors 4 larger ones 3 black striped smaller ones There is a soldered in ceramic 3 15A internal fuse F1 Disassemble unsolder and replace if needed J remove front plastic bezel there are top and bottom plastic snaps K lubricate the three X axis rails with Teflon with several drops of lube Use a flat bladed screwdriver and su...

Page 6: ...the mechanism back and forth from end to end When satisfactory return the carriage to a central position If a medical syringe is available it may be easier but the Teflon may settle out and the vapor pressure of carrier vehicle may cause it to squirt out on unintended places N On the bottom of the mechanism there is a spring loaded brown bronze clip spring covering the Autofocus Z axis stepper mot...

Page 7: ... scanning assy is moved back and forth Time to test the bare mechanism under live power be careful to not touch power supply parts which may have dangerous high voltage Plug in power cord and turn on power switch S watch the shaft of Z axis Autofocus stepper motor during the beginning part of the boot process it must rotate back and forth The carriage can be manually moved out of mid range in both...

Page 8: ... in elongated sliding slots Fasten in the mid position IV Warranty and Limits of Liability This procedure is provided on a best efforts basis and there is no warranty whatsoever herein and is provided AS IS It requires a high level of technical competence and mechanical dexterity such as an A Certification Computer Technician plus 3 years experience or equivalent and demonstrated ability to handle...

Page 9: ... chip Firmware 30 soldered in main chips Nikon EI 104 vE17 main controller Hitachi HD6413003F16 H8 3003 microcontroller Analog Devices AD9220AR 12 bit A D converter Analog Devices AD7528JR 8 bit D A converter LSI Symbios 53CF92A SCSI controller WinBond W24257AJ 15ns memory x8 JRC 7809A voltage regulator high quality capacitors are metal encased CCD Signal Processing board on sensor assy P N F17045...

Reviews: