P-400R Digital Microscope Instructions
84
9
T
roubleshoot
ing
9.2
About the Image Display
Symptom
Possible cause
Action
Ref.
A lens cover is
attached.
Remove the lens cover and direct the lens
to the object.
26
The illuminator is
turned off.
Turn on the illuminator so that the object
is illuminated adequately.
28
Illumination intensity
is not sufficient.
Adjust the illuminator to adequate
intensity.
29
The exposure setting
is not adequate.
Adjust the exposure so that an image with
proper brightness can be displayed.
32
36
An external monitor is
connected.
Disconnect the P-CV video cable from the
P-400R. Operate the power switch of the
motorized focusing stand.
15
The P-400R is
attached to the
motorized focusing
stand.
When the P-400R is mounted on the
motorized focusing stand, the LCD
monitor of the P-400R is turned off.
See the
ShuttlePix
Digital
Microscope
Instructions.
P-RELS Release is
connected to the
P-400R.
Turn off the video output signal when the
P-RELS Release is connected, or
disconnect it.
42
76
No image is
displayed in
the LCD
monitor.
No SD card is
inserted.
Insert an SD card.
11
The cable is not
connected correctly.
Use the optional P-CV video cable to
connect the P-400R to the external
monitor.
15
The power of the
external monitor is off.
Turn on the power of the external monitor.
15
The input of the
external monitor is not
switched correctly.
Switch to the external input.
15
No image is
displayed in
the external
monitor.
The video output
signal format is
different.
Set the P-400R video output signal format
appropriate for the external monitor to be
used. You can select either NTSC or PAL
as the video output signal format.
76