Assembly

Required tools:

Hexagonal wrenches (large / small)(supplied with the teaching head
stand)

Hexagonal screwdriver (supplied with the teaching head stand)

Hexagonal screwdriver (for mounting the pointer unit: M3)

Minus screwdriver

1

Attach the pillar to the base.

Use the large hexagonal wrench to securely
tighten the three screws.

2

Attach the focusing mount to the pillar.

Slide on the focusing mount along the
dovetail mount on the pillar until it stops at
the limit pin.  Tighten two clamp screws with
a minus screwdriver.

3

Level the base.

Use the small hexagonal wrench to turn the
base leveling screw to stabilize the base.

Large  hexagonal 
wrench

Limit pin

Clamp 
screws

Small hexagonal 
wrench

Base

Base leveling screw

The pillar should
be attached here.

Summary of Contents for P-THS

Page 1: ...P THSS Teaching Head P THS Teaching Head Stand P THSS ティーチングヘッド P THS ティーチングヘッド用スタンド Instructions 使用説明書 ...

Page 2: ......

Page 3: ...d permission The contents of this manual are subject to change without notice Although every effort has been made to ensure the accuracy of this manual if you note any points that are unclear or incorrect contact your nearest Nikon representative Some of the products described in this manual may not be included in the set you have purchased Be sure to read the instruction manual for any other prod...

Page 4: ...piece tube clamp Eyepiece tube mount for the main observer Eyepiece tube clamp Pointer mount Screw hole for attaching the epi arm Base leveling screw Vinyl sheet Vertical movement clamp screw Vertical movement handle Pillar Pillar fixing screws Focusing mount clamp screws Focusing mount Focus knob Teaching head clamp screws ...

Page 5: ...ge hexagonal wrench to securely tighten the three screws 2 Attach the focusing mount to the pillar Slide on the focusing mount along the dovetail mount on the pillar until it stops at the limit pin Tighten two clamp screws with a minus screwdriver 3 Level the base Use the small hexagonal wrench to turn the base leveling screw to stabilize the base Large hexagonal wrench Limit pin Clamp screws Smal...

Page 6: ...d and while pushing the zooming body toward the front tighten the clamp screw with a hexagonal screwdriver Be sure to hold the zooming body with your hand until it is securely clamped on to the teaching head 6 Attach the eyepiece tubes for the main and secondary observers Matching the groove on the bottom surface of the eyepiece tube with the pin on the mount attach the eyepiece tube on the teachi...

Page 7: ...When attaching 10X eyepieces to the eyepiece sleeves do not hold the diopter adjustment rings but hold their rubber covers to avoid damage to the eyepiece Also note that because of these covers you cannot check with your eyes that the eyepieces are securely inserted in place Feel through your fingers that they are securely inserted in place 8 Screw on an objective to the zooming body ...

Page 8: ...precisely focus on the sample See the instruction manual supplied with the stereoscopic microscope for the working distances of the objectives Adjust the diopter before observation See the instruction manual supplied with the microscope for the details on the diopter adjustment Avoid the following operations since they will damage the instrument Turning the vertical movement handle without looseni...

Page 9: ...sing a hexagonal screwdriver Connect the Y THAC THAC adapter to the side of the pointer unit Connect the AC adapter to the wall outlet Lamphouse cover Before replacing the lamp turn off the lamp with the arrow brightness control turn counterclockwise until it clicks unplug the AC adapter from the wall outlet and wait till the lamp cools down Arrow rotation dial Turn to change the arrow direction A...

Page 10: ...g the grooves with the protrusions push in and twist clockwise till the limit 3 Specifications Y THP TH pointer unit Type name THP Input voltage DC 9V 0 5A max Lamp for arrow illumination 6V 0 2A tungsten lamp average lifetime 1000 hrs Arrow color Green or orange can be selected Arrow illumination voltage 0 to 6V can be controlled by brightness control which also functions as ON OFF switch Conform...

Page 11: ...en illuminators to the epi arms See the instruction manual supplied with the stereoscopic microscope for the use of halogen illuminators 4 P ICI2 coaxial episcopic illuminator The P ICI2 coaxial episcopic illuminator can be attached between the teaching head and the zooming body See the instruction manual supplied with the stereoscopic microscope for the use of P ICI2 coaxial episcopic illuminator...

Page 12: ...C BD115 diascopic bright dark field stand 115 C BD230 diascopic bright dark field stand 230 Place the diascopic stand on the base matching their rubber feet with the dents φ13 depth 0 7mm on the base See the instruction manual supplied with the stand for their usage 6 Vinyl sheet Lay the vinyl sheet on the base as necessary such as when you do not want your sample to get scratches etc ...

Page 13: ... 0 61 4 0 17 4 Plan Apo 0 5X 0 57 9 0 13 9 Cannot be used Cannot be used Plan Apo 1X 0 135 0 91 0 49 9 0 5 9 ED Plan 1 5X 0 143 5 0 99 5 0 58 4 0 14 4 ED Plan 2X 0 141 Cannot be used 0 55 9 Cannot be used SMZ800 Achro 0 5X 0 69 5 0 25 5 Cannot be used Cannot be used Plan 1X 0 142 5 0 98 5 0 57 4 0 13 4 Plan 1X ERGO 0 159 5 0 115 5 0 74 4 0 30 4 Plan Apo 0 5X 0 70 9 0 26 9 Cannot be used Cannot be ...

Page 14: ...い致します 本書の内容の一部 または全部を無断で転記することは禁止されて います 本書の内容については 将来予告なしに変更することがあります 本書の内容については万全を期しておりますが 万一不可解な点や 誤り お気付きの点がございましたら ご購入先にご連絡ください ますようお願い致します 本書に掲載されている製品の中には ご購入頂いたセットに含まれ ないものがある場合もあります 実体顕微鏡や同軸落射照明装置など 同時にお使いになる製品の使 用説明書も併せてお読みになり そこに書かれた指示は必ずお守り ください P THS ティーチングヘッド用スタンドは SMZ1500 SMZ1000 および SMZ800 専用です ...

Page 15: ... 2 I 各部の名称 ベース P THSティーチング ヘッ ド用スタン ド P THSS ティーチングヘッ ド 従観察側鏡筒受け 鏡筒クランプねじ 主観察側鏡筒受け 鏡筒クランプねじ ポインタユニッ ト 取付け部 エピアーム取付け穴 ベース高さ調節ねじ 塩ビシー ト 上下動クランプねじ 上下動ノブ 上下動部 上下動部固定ねじ 焦準装置クランプねじ 焦準装置 焦準ノブ ティーチングヘッ ド クランプねじ ...

Page 16: ...付け用 M3 マイナスドライバ 1 上下動部をベースに取り付けます 付属のねじ 3 本を使って 六角レンチ 大 でしっ かりと固定してください 2 焦準装置を上下動部に取り付けます 上下動部のアリに 焦準装置を制限ピンに当たるま で差し込みます 2 本のクランプねじをマイナスド ライバでしっかりと締め 固定します 3 ベースががたつかないよう調整します ベース高さ調整ねじを六角レンチ 小 で回して が たつかないよう調整します 六角レンチ 大 制限ピン クランプ ねじ 六角レンチ 小 ベース 上下動部 取付け位置 ベース高さ調整ねじ ...

Page 17: ...プねじを マイナスドライバでしっかりと締め 固定します 5 実体顕微鏡を取り付けます 主観察側の底面にある丸アリに 顕微鏡のズーム本 体をはめ込み 手前側に押しつけた状態で 六角ドラ イバでクランプねじを締めます ズーム本体の固定 が終わるまで しっかりと顕微鏡を支えておいてく ださい 6 主観察側 従観察側の鏡筒を取り付けます 鏡筒底部にある溝と 鏡筒受けのピンを合わせて鏡 筒を取り付け 六角ドライバでクランプねじを締め て固定します 制限ピン クランプねじ 2ヶ所 顕微鏡を 手前側に 押しつけ ながら クランプ する ピン 溝 ...

Page 18: ...J 5 取付け方 II 7 接眼レンズを取り付けます 鏡筒のスリーブ端面に当たるまでしっかりと差し込みます 注 10 接眼レンズにはゴムカバーが付いていて 端面が見えませんので感触で確 かめてください 視度調節リングを持って押し込むと故障の原因となりますの で ゴムカバーの部分を持って差し込んでください 8 顕微鏡のズーム本体に対物レンズをねじ込みます ...

Page 19: ...しくなるよう調節し 上下 動クランプねじを締めて固定します 接眼レンズをのぞきながら 焦準ノブを回して 試料に正確にピントを合わせます 対物レンズの作動距離については 実体顕微鏡に付属の使用説明書をご覧くださ い 視度調節を行ってからお使いください 視度調節については 実体顕微鏡に付属の 使用説明書をご覧ください 故障の原因となりますので 次のような操作を行わないでください 上下動クランプねじ締めたまま 上下動ノブを回す 左右の焦準ノブをそれぞれ逆の方向に回す 焦準ノブを制限に当たった後も更に回転させる 上下動クランプねじ 焦準ノブ 上下動ノブ 上がる 作動距離 ...

Page 20: ... 付け M3の六角ドライバを使って 4本のねじを締 め 固定します ポインタユニットの側面にY THAC THACアダプタを接続します 最後に ACアダプタ のプラグをコンセントに接続します III ランプカバー ランプ交換は ランプが冷えてから 行ってください ランプ交換時は 矢 印調光ボリュームを反時計方向にカ チッと音がするまで回して矢印を消 灯し ACアダプタをコンセントから 抜いてください 矢印回転ダイヤル 回すと 矢印の向きが 変わります 矢印調光ボリューム ON OFFスイッチ兼用 時計方向に回すと矢印が明るく なります 反時計方向にカチッ と音がするまで回すと電源が OFFになり 矢印が消灯します 矢印移動つまみ 矢印の場所が移動します 矢 印が視野の外にある場合は 矢 印移動つまみを操作して 矢印 を視野内に戻してください 矢印色切替えダイヤル Gの方向に回すと緑 Rの方...

Page 21: ...カバーを取り付けます 溝と突起を合 わせて差し込み 時計方向に制限に当たるまで 回転させます 3 仕様 Y THP TH ポインタユニット 型式名称 THP 入力電圧 DC 9V 0 5A 最大 矢印照明用電球 6V 0 2A タングステンランプ 平均寿命 1000 時間 矢印色 緑 オレンジ 切替可能 矢印調光電圧 0 6V ON OFF 兼用調光ボリュームによる 適合規格 本製品は EU の低電圧指令を満足しています 本製品は EU の EMC 指令を満足しています Y THAC THAC アダプタ 型式名称 SA 091A0F 10 入力電圧 AC 100 250V 0 3A 最大 出力電圧 DC 9V 1A 最大 適合規格 本製品は UL 規格を取得しています 本製品は CSA を満足しています 本製品は EU の低電圧指令を満足しています 本製品は EU の EMC 指令を満足し...

Page 22: ...同軸落射照明装置 3 エピアーム 焦準装置の底面に エピアームを取り付けるための ねじ穴が 2 箇所あります ここにエピアームを取り 付け エピ照明装置を使用することができます エピ 照明装置の使い方については 実体顕微鏡に付属の 使用説明書をご覧ください 4 同軸落射照明装置 実体顕微鏡のズーム本体とティーチングヘッドの間 に同軸落射照明装置を取り付けることができます 同軸落射照明装置の使い方については 実体顕微鏡 に付属の使用説明書をご覧ください ...

Page 23: ...ンドシリーズ 高級透過照明スタンド C DSD115 C DSD230 透過照明スタンド C DSS115 C DSS230 明暗視野照明スタンド C BD115 C BD230 ベースのくぼみ Ф 13 深さ 0 7mm にスタンド のゴム足が入るように スタンドをベースの上に載 せます 各スタンドの使い方については スタンドに 付属の使用説明書をご覧ください 6 塩ビシート 試料に傷を付けたくない場合など 必要に応じて塩ビシートをベース上に敷いてくだ さい ...

Page 24: ...HR Plan Apo 1 0 124 2 0 80 2 0 39 1 使用不可 HR Plan Apo 1 6 0 132 9 使用不可 0 47 8 使用不可 SMZ1000 Achro 0 5 0 56 5 0 12 5 使用不可 使用不可 Plan 1 0 129 5 0 85 5 0 44 4 使用不可 Plan 1 ERGO 0 146 5 0 102 5 0 61 4 0 17 4 Plan Apo 0 5 0 57 9 0 13 9 使用不可 使用不可 Plan Apo 1 0 135 0 91 0 49 9 0 5 9 ED Plan 1 5 0 143 5 0 99 5 0 58 4 0 14 4 ED Plan 2 0 141 使用不可 0 55 9 使用不可 SMZ800 Achro 0 5 0 69 5 0 25 5 使用不可 使用不可 Plan 1 0 142 ...

Page 25: ......

Page 26: ...千代田区丸の内 3 2 3 富士ビル 112 0004 東京都文京区後楽 1 4 25 日教販ビル 8F FAX 03 5802 0225 工業顕微鏡部 電話 03 5802 0214 測定機部 電話 03 5802 0215 販売部エンコーダグループ 電話 03 5802 0216 販売部産業機器グループ 電話 03 5802 0217 バイオメディカル部 電話 03 5802 0218 眼鏡機器部 電話 03 5802 0223 001 0011 札幌市北区北 11 条西 4 1 40 すばるビル 電話 011 746 9381 FAX 011 746 9398 980 0014 仙台市青葉区本町 2 15 1 ルナール仙台ビル 電話 022 227 1298 FAX 022 227 0331 460 0002 名古屋市中区丸の内 3 20 17 中外東京海上ビル 電話 052 954...

Reviews: