Chapter 5
Assembling
78
2
Attach the intelligent nosepiece or focus mount adapter.
Attach the intelligent nosepiece or focus mount adapter to the focus unit.
Both the intelligent nosepiece and focus mount adapter are attached to the focus unit in the same manner.
The following figure shows an example of attaching the intelligent nosepiece to the P2-FU Focus Unit.
(1) Remove the two rubber caps from the fixing bolt
holes at the right rear of the intelligent nosepiece (as
viewed from the front). Insert the hex driver (nominal
designation: 4) and completely loosen the two
internal M5 hexagon socket head bolts.
(2) Drop and fit the dovetail groove on the rear side of
the intelligent nosepiece onto the dovetail of the
focus unit mount from the top.
Attaching the intelligent nosepiece
(3) Firmly tighten the two fixing bolts using a hex driver
(nominal designation: 4). After tightening the bolts,
put the rubber caps back onto the bolt holes.
Firmly tightening the bolts
(4) The cover on the left side (as viewed from the front)
of the intelligent nosepiece is a sliding cover.
Removing this cover at this point will facilitate the
cable connection and routing in “[4] Connect the
signal cables.”
Remove the rubber caps from the two fixing bolt
holes at the bottom of the cover, loosen and remove
the two internal M5 hexagon socket head bolts using
a hex driver (nominal designation: 4), and pull the
cover up to remove it.
Summary of Contents for SMZ18
Page 1: ... M608EN01 Research Stereo Microscope SMZ25 SMZ18 Instruction Manual M608 E 13 5 NF 1 ...
Page 2: ......
Page 20: ...Chapter 1 Components 6 ...
Page 32: ...Chapter 2 Microscopy Procedures 18 ...
Page 84: ...Chapter 4 Functions and Operations of the Devices 70 ...
Page 114: ...Chapter 6 Troubleshooting 100 ...
Page 118: ...Chapter 7 Maintenance and Storage 104 ...
Page 119: ...105 Chapter 8 Specifications ...