VRA51701-R.3607.A
-
A39
・
-
LS-9000
・
Scanning position accuracy
Select "Run" on the main menu, then choose "Mechanical Precision Test" and "Precision the Scanning Position" from
the pulldown menus.
In case of NG: Readjust the alignment mark.
Inspection result on scanning position accuracy
・
Main/Sub-scan perpendicularity
Select "Run" on the main menu, then choose "Mechanical Precision Test" and "Main/Sub-scan Perpendicularity" from
the pulldown menus.
In case of NG: Inspect the stage carriage and CCD.
Inspection result on
Main/Sub-scan perpendicularity