31
KSKY/02/04.09.2019
4)
Install the lid which filter is installed to its place.
Don’t reuse the used filter by cleaning again
.
Just use the filters which are produced by manufacturer otherwise the device
can be harmed
görebilir.
Never use device without filter.
➢
Because of using frequently and disinfection processes of nebulizers and
accessories plastic parts exposure fray. This can change the aerosol specialities
which is coming from medication cup and effects the inhaling negatively. For this
reason it is recomended to change medication cup, masks, air-tube and
mouthpiece after 1 year.
➢
Please protect the device, mediction cup and accessories against hard impulse.
Otherwise this can couse damage and device will become useless
➢
Clean the device and its parts before every usage.
➢
Do not extract, fix or any change on device otherwise this can cause electric
shock, fire or leckage.
➢
Never use the device if it is damaged or fall in to water.
If any problem occures while using the device first check the troubleshooting.
If problem continuous apply to the authorised service.
9.
Device’s Technical Datas
➢
The device is manufactured in compliance with EN 13544-1:2007+A1:2009
standards for airway treatment devices
➢
The nebulisation system datas are given in the table in down side.
Nebulization
Control Valve
Closed
Nebulization
Control Valve
Opened
MMAD (Particle Size)
3,19 μm
3,48 μm
Nebulization Rate
0,2 ml/min
0,4ml/min
GSD (Geometric Standard Deviation)
2,32
2,20
Particle Size Under 5 μm
Over %70
Over %68
➢
The horizontal axis is the particle size value, the value is logarithmic
distribution;The left vertical axis is the cumulative percentage of the volume,
corresponding to the rising trend of the curve