21
Note:
If Meter does not beep or begin countdown
soon after placing Sample Tip to Control drop,
discard Test Strip. Repeat test with new Test
Strip. If problem
persists, see
Troubleshooting
.
11. Compare result to Control range
printed on Test Strip vial label.
If result is in range, System
can be used for testing blood.
If result does not fall within
range, repeat test using a new
Test Strip.
Note:
Control Test result shows the
Control Symbol in the Display.
Note:
When using the meter for the
first time, check the unit of
measure in the meter Display after performing
a control or blood test. Result must display
in the correct unit of measure (mg/dL). If the
result does not display in the correct unit of
measure,
DO NOT
use the meter and
DO NOT
use the meter result to determine treatment.
Caution!
If Control Test result is outside range, test
again. If result is still outside range, System
should not be used for testing blood. Call
for assistance (see cover for phone number).
12. Remove Strip from Meter and discard. Meter
turns off. Recap Control bottle tightly.