GETTING STARTED
The TRUEyou Meter comes with pre-set time, date, Ketone
Test Alert and all Testing Reminders off. Before using
Meter for the first time or after bat tery is changed, check
to make sure all of the Meter pre-set functions are correct.
Update any setting as needed (see
Meter Set Up
).
The Meter turns on when a Test Strip is inserted into the
Test Port or when “ S ” Button is pressed and released (see
Memory
and
Meter Set Up
). Meter turns off when the Test
Strip is released from the Meter, “ S ” Button is pressed and
held for 20 seconds, or after 2 minutes of non-use.
Always check your supplies before using.
• Check Meter for damage (cracked Display, missing
buttons, etc.). If damage is seen, do not use Meter. Call
for assistance.
• Check Test Strip vial for damage (cracked vial, broken
vial, etc.). Discard damaged vial and all contents (Test
Strips) and use a new vial of Test Strips for testing.
• Write date first opened on Test Strip vial. Check Use By
Dates (written and printed) before using any Test Strips
from the vial. Do not use if 4months after first opening
(written date) or if printed Use By Date has passed.
• Check Control Solution bottle for any leaks or broken
cap. Discard bottle and open a new one for testing.
• Write date first opened on Control Solution bottle
label. Check Use By Dates (written and printed) before
using. Do not use if 3 months after first opening
(written date) or if printed Use By Date has passed.
14