11. Compare result to
Control Test Range
printed on Test Strip
vial label for Control
Solution Level you
are using.
If result is in range,
System can be used for
testing blood. If result
does not fall within range,
repeat test using a new
Test Strip.
If result is still outside range after a second
Control Test, System should not be used for
testing blood. Call for assistance (see cover for
phone number).
12. After result is shown, Strip Release
Button flashes. Hold Meter with Test
Strip pointing down. Press Strip Release
Button to release and discard Test Strip
in appropriate container. Meter turns off.
Recap Control Solution bottle tightly.
Note:
Removing Test Strip before result displays cancels
the test. An error message appears and the result
is not stored in Memory. Retest with a new Test
Strip and do not remove before result is displayed.
19
Control Symbol
Test Strip
Vial Label
Control Solution
Bottle Label