For Healthcare Professionals
99.3% of healthcare professional (HCP) TRUEyou fingertip values fell within
0.83 mmol/L of the YSI results at glucose levels < 5.55 mmol/L and within
15% at glucose levels > 5.55 mmol/L.
Fingertip Samples (HCP vs. YSI) for glucose concentrations <5.55 mmol/L
0.28 mmol/L 0.56 mmol/L 0.83 mmol/L
74/180 (41%)
155/180 (86%)
178/180 (98.8%)
Fingertip Samples (HCP vs. YSI) for glucose concentrations >5.55 mmol/L
5%
10%
15%
272/420 (71%)
395/420 (94%)
418/420 (99.5%)
Fingertip Samples for glucose concentrations between 1.1-33.3 mmol/L
0.83 mmol/L or + 15%
596/600 (99.3%)
Parkes Error Grid:
100% of individual fingertip glucose measured values
performed by healthcare professionals fell within Zone A of the Parkes Error
Grid (PEG).