For Consumers
99.0% of user TRUEyou fingertip values fell within 0.83 mmol/L of the YSI results at
glucose levels < 5.55 mmol/L and within 15% at glucose levels > 5.55 mmol/L.
Fingertip Samples (user vs. YSI) for glucose concentrations <5.55 mmol/L
0.28 mmol/L 0.56 mmol/L 0.83 mmol/L
11/23 (48%)
21/23 (91%)
23/23 (100%)
Fingertip Samples (user vs. YSI) for glucose concentrations >5.55 mmol/L
5%
10%
15%
38/77 (49%)
66/77 (86%)
76/77 (99%)
Fingertip Samples for glucose concentrations between 1.1-33.3 mmol/L
0.83 mmol/L or + 15%
99/100 (99%)
Parkes Error Grid:
100% of individual fingertip glucose measured values
performed by users fell within Zone A of the Parkes Error Grid (PEG).