EN-13
Disposing Used Test Strip and Lancet
To remove the used test strip, simply push the
Test Strip Ejector
button upward to
eject the used test strip. The device will automatically turn off after the test strip is
removed.
To remove the used lancet, remove the lancet from the lancing device after you have
finished testing. Discard your used strip and lancet properly in a puncture-resistant
container.
Important!
The used lancet and test strip may be biohazards. Please consult your health-care
provider for proper disposal which complies with your local regulations.
7. Control Solution Testing
Important!
Please note there are two main types of 4SURE control solution available;
1. 4SURE blood glucose control solution (Level 1, 2, 3). This solution contains a known
amount of glucose that reacts with the 4SURE blood glucose test strips and is used to
ensure your meter and test strips are working correctly.
2. 4SURE β-Ketone control solution (level 1, 2). This solution contains a known amount
of β-Ketone that reacts with the 4SURE β-Ketone test strips and is used to ensure
your meter and test strips are working correctly.
Only use 4SURE blood glucose control solution to perform a blood glucose control
tests. Only use 4SURE β-Ketone control solution to perform a β-Ketone control test.
Control solution is not provided within your meter starter kit but is available from
customer service. Tel: 0800 08 588 08.
Advice on when to perform a Control Solution Test:
9
You suspect the device or test strips are not working properly.
9
Your blood glucose or β-Ketone test results are not consistent with how you feel, or
if you think the results are not accurate.
9
You have dropped or think you may have damaged the device.
To perform a blood glucose control solution test, do the following:
1. Insert the blood glucose test strip into the test slot of
the device. Wait for the device to display the test strip “
“ and blood drop “ “.