6 Installation
NivuLink Micro - Rev. 02/13.11.2018
page 31
Close the enclosure
Procedure
Important Note
Particles and pieces of dirt can damage the seal inside the cover.
−
Before you close the enclosure, remove any particles and pieces of dirt.
The manufacturer will not accept responsibility for damage to the device caused by
leaky or damaged seals, or if the cover is incorrectly closed.
1. If necessary, remove any particles and dirt from the seal inside the cov-
er
2. Close the cover (
3
)
CAUTION
Device damage caused by dirt or water
If dirt or water enters the terminal compartment, the protection class for the device
is no longer valid.
−
Close the terminal compartment with the cover and the two screws so that
no water or dirt can enter.
3. Make sure the cover is correctly closed and that no particles are caught
between enclosure und cover.
CAUTION
Risk of damage to device
If the screws are tightened with the incorrect torque, the protection class for the
device is no longer valid.
−
Tighten the screws with 1.50 Nm torque.
4. Insert screws (
2
) and tighten with 1.50 Nm torque
5. Tightly press lock (
1
) against the enclosure until it snaps into place.
The lock is closed