CAUTION
Damage due to weather influences
The instrument will be damaged as soon as water is leaking into the enclosure.
−
In case of rainfall or any other poor weather conditions causing precipitation
or potential ingress of water from above the NivuLinkMicro shall be protect-
ed appropriately from water leaking into the enclosure
−
Make sure to protect the enclosure against ingress of water
−
Do not operate NivuLinkMicro out in the field with the cover opened
6.6.2
Notes on cabling
The NivuLink Micro is shipped with the following cable glands:
•
2 x thread adapter M16 x 1.5
•
1 x plug for thread adapter M16
Cables with the following external cross-sections can be used with the shipped
thread adapters:
•
M16 x 1.5
5 – 9 mm
Important Note
The protection rating IP68 is not guaranteed if cables with diameters greater than
these values are used.
Important Note
Before initial startup:
Close any unused cable glands with a suitable plug. Otherwise, the protection rat-
ing for the NivuLink Micro is not guaranteed and the manufacturer's warranty is