Ext. temperature
sensor
Off
---
Measurement channel deactivated
On
Hold
Maintain the last valid measurement value for x
measurement cycles
Off
Function deactivated
1-5
Number of measurement cycles for which
the measured value is maintained until the
error value is issued
On
In the event of an error, the last valid
measurement value is maintained until a new
valid measurement value is present.
9.2.1.3.3 Alarms
"Digital" or "Digital LP"
mode
WA
A "high" at the universal input initiates a "warning"
AL
A "high" at the universal input initiates an "alarm"
SW
A "high" at the universal input initiates a "fault warning"
SA
A "high" at the universal input initiates a "fault alarm"
All other modes
Warning
Value low
A warning is initiated if the measurement value drops to or
below this value.
Value high
A warning is initiated if the measurement value meets or
exceeds this value.
Alarm
Value low
An alarm is initiated if the measurement value drops to or
below this value.
Value high
An alarm is initiated if the measurement value meets or
exceeds this value.
Hyst %
Hysteresis for all-clear (e.g. hyst = 5%, alarm or warning at 100 -> all-
clear at 95)
9.2.1.3.4 Trigger
"Digital" or "Digital LP"
mode
QU
Fastrecording (recording cycle = recording cycle / factor)
SL
Slow recording (recording cycle = recording cycle * factor)
MS
Start measuring cycle immediately
XM
Initiate transmission
ON
Activate online mode
Edge
Selection of the edge at which the trigger should be initiated
Rising
Rising edge initiates trigger.
Falling
Falling edge initiates trigger.
Both
Both edges initiate trigger.
52 of 73
NivuLog Easy - Rev. 05 as of 02.07.2012