Read/write registers
82
Figure 48
Setting the modulation range parameter
Setting the
wavelength
modulation source
To configure the
for the wavelength modulation option, set
the following bits of register 0x31:
•
Bit 2 – sets external wavelength modulation where: 0 disabled/1 enabled.
•
Bit 4 – sets external wavelength modulation where: 0 disabled/1 enabled.
Figure 49
Setting the wavelength modulation signal source
The laser supports three different signal source combinations to modulate the
wavelength. Below are some example 0x31 register settings for all three source
settings.
•
External
•
0x31 set to 0504 – external source - wide range
•
0x31 set to 0506 – external source - narrow range
•
Internal
•
0x31 set to 0510 – internal (generator) source - wide range