103
Connecting the optical output
Output trigger
When the trigger feature is set as an output, the trigger pin should be connected
to an external pull-up resistor. The diagram
shows the trigger output
circuit.
Figure 70
Trigger output circuit
The high and low voltage levels are defined by the following equations:
V
high
= V
external
V
low
= V
external
x 22k / (22k + R)
Connecting the optical output
Before connecting the optical output connectors, ensure to check the connector
tips using a fiber microscope. Check for any deformities, damage, residue or
other contaminants at the optical tip of the connector. Either clean the connector
or contact NKT Photonics support if replacement is necessary.