17
2 Installation
Mounting the
CONNECT
Note:
Ensure to mount and fasten the SuperK CONNECT to your accessory before
turning on your laser.
Warning:
The input power to the fiber delivery system (FDS) should always be lim-
ited to below 500 mW.
Note:
Ensure to select the correct SuperK CONNECT model (VIS or IR) that match-
es with your accessory output port.
To mount a SuperK CONNECT to a laser accessory, follow the steps in
Procedure 1
Mounting a SuperK CONNECT to an optical output port.
1.
Turn off the SuperK laser and ideally, disconnect power from it.
2. Locate the optical output port on your accessory to which the FDS will be
attached.
3. If the port has a cover plate, remove it.
4. Using four M3 x 6mm machine screws, select a vertical or horizontal
orientation (see
“VIS or IR SuperK CONNECT orientation”
below) and fasten
the SuperK CONNECT to the port.
5. Install the FD following the steps in
.
VIS or IR SuperK
CONNECT
orientation
The correct orientation of a SuperK CONNECT with respect to the optical output
port of the CONNECT depends on whether or not the output is coupled to a
polarization maintaining fiber (PM).
•
For non-PM fiber
– the orientation of the CONNECT is not critical and can be
oriented in any direction. However, ensure to consider access to the SuperK
CONNECT alignment screws and the input coupler locking screw in the final
installation.
•
For PM fiber
– the SuperK CONNECT must be mounted in the “horizontal”
orientation in order to couple the polarization of the visible beam to the slow-
axis of the fiber.
shows examples of a SuperK CONNECT mounted in both horizontal and
vertical mounting orientations.