35
A Specifications
Table 2
FD Fiber wavelength and transmission characteristics
FD fiber type
Cut-off λ
% Transmission
Typical peak % transmission
FD1-PM
425±25 nm
> 60% from 425 to 775 nm
75% @ 650 nm
FD2
450±25 nm
> 60% from 450 to 775 nm
80% @ 650 nm
FD3
630±20 nm
> 60% from 630 to 1100 nm
80% @ 800 nm
FD3-PM
580±40 nm
> 65% from 580 to 950 nm
85% @ 650 nm
FD4
730±30 nm
> 70% from 730 to 1150 nm
80% @ 950 nm
FD4-PM
710±60 nm
> 65% from 710 to 1100 nm
70% @ 900 nm
FD5
930±40 nm
> 50% from 930 to 1550 nm
65% @ 1350 nm
FD5-PM
900±70 nm
> 50% from 900 to 1500 nm
60% @ 1200 nm
FD6
1260±40 nm
> 30% from 1260 to 1900 nm
40% @ 1650 nm
FD6-PM
1200±70 nm
> 30% from 1200 to 1900 nm
35% @ 1700 nm
FD7*
< 400 nm
> 70% from 450 to 1050 nm
80% @ 750 nm
FD7-PM*
< 400 nm
> 70% from 450 to 950 nm
80% @ 600 nm
FD8*
< 800 nm
> 40% from 800 to 1700 nm
55% @ 1300 nm
FD9*
< 500 nm
> 50% from 500 to 1300 nm
60% @ 900 nm
FD10**
NA
> 80% from 500 to 1100 nm
90% @ 750 nm