It is important to thoroughly review this 
manual and test your new duplicator prior 
to loading with duplicating material.

I. Receiving & Unpacking

We have made every effort to test your
equipment and inspect it for damages prior
to packing for shipment. However, once the
equipment leaves our factory, the shipping
company accepts the equipment and the
responsibility becomes theirs for its safe and
undamaged delivery.

When you unpack your shipment, keep all
the packing materials until you are confident
that the equipment is in good working 
condition and functions properly. If there 
is any concealed damage, the shipping 
company may need to examine the container
when a claim is filed.

If there is damage, the carrier should be 
contacted as soon as possible. The claim 
for damages must be filed with the carrier.
We are not in a position to assist you with
replacement, repair or warranty issues until
the carrier has completed their investigation.

If no damage has been found, place the
duplicator on a level bench where it will be
used. It is best to locate the unit adjacent 
to the area where the duplicate models 
will be cooled. Do not block the exhaust fan.

II. Electrical Connection

The Colloid-Saver 3.5 duplicator requires
only a standard 115V–60 Hz outlet to 
operate. Before plugging-in the duplicator,
be sure to identify the electrical specifica-
tions of the outlet. A dedicated circuit
should be used as the Colloid-Saver 3.5
duplicator draws about 12 amps at 115 volts.
As with any electrical device, take extra care
when handling liquids around electrical 

connections to avoid accidental shock.  
It is important to carefully read and follow
the instructions in the next section 
prior to actually plugging-in the duplicator.

III. Testing & Inspection

This is a very important step in the installa-
tion of your new Colloid-Saver 3.5 duplicator.
Even though your unit was fully tested and
inspected before leaving our factory, it may
have suffered damage during transit. For this
reason the following testing and inspection
steps are very important.

Note: The parts list located in the back of this 
manual will help in the proper identification of the
items described below.

1. DO NOT insert the power cord. 
Remove the rear panel and inspect 
internal components visually. Look for 
loose connections and parts.

2. Replace rear panel after inspection.

3. Remove the top cover by rotating it 
clockwise and check inside the tank to 
be sure there are no interferences with 
the stirrer. Make sure stirrer knob is tight.

4. Replace top cover and rotate it 
counter-clockwise to engage it with the 
locking pins in the tank. Make sure cover
handle is facing forward. 

5. Check to be sure the duplicator’s 
identification plate voltage agrees with 
the electrical voltage of the outlet. If unsure, 
test voltage or contact a qualified electrician.
If these two items are compatible, plug
duplicator into the electrical outlet.

Note: The Colloid-Saver 3.5 duplicator (115V) 
draws about 12 amps and should be on its own 
dedicated circuit.

3

Summary of Contents for Ticonium Colloid-Saver 3.5

Page 1: ...for precision temperature control SmartCycle stirring cycle activated only when needed minimizing colloid breakdown High volume no drip no clog dispensing nozzle Pre set programs for virtually any col...

Page 2: ...ty If you have any questions regarding CMP s quality line of dental laboratory equipment and supplies please call toll free 800 833 2343 or 518 434 3147 Fax 518 434 1288 between the hours of 8 00 AM a...

Page 3: ...ed as the Colloid Saver 3 5 duplicator draws about 12 amps at 115 volts As with any electrical device take extra care when handling liquids around electrical connections to avoid accidental shock It i...

Page 4: ...ke sure to add any liquid that may be in the bottom of colloid container The liquid contains ingredients which extend the life of the colloid Replace top cover and rotate it counter clockwise to engag...

Page 5: ...at the very bottom of the tank simply allow it to cool then peel it out Preventive Maintenance Cleaning 1 Clean Colloid Saver 3 5 duplicator approximately once a month or when new duplicating material...

Page 6: ...rofile settings below To select a specific profile the duplicator must be in Standby Mode power on but no heating lights illu minated Press the SET button twice The display will show the first option...

Page 7: ...life VII Troubleshooting 1 Colloid Saver 3 5 duplicator will not turn on A Check wall outlet for power B Check if power cord is connected to duplicator C Check if the main power supply is turned to th...

Page 8: ...P37 2 Precision Laser 40811P48 3 LED Light 40811P38 4 Power Entry Module 40811P20 5 Cover Lock 40811P44 6 Poly Cover 40811P1 7 Cover Interlock 40811P45 8 Cover Handle 40811P43 9 Power Cord Assembly 40...

Page 9: ...ART NUMBER 1 USB Port Connector 40811P39 2 1 Amp Motor Circuit Breaker 65383P7 3 15 Amp Main Breaker 65383P6 4 Power Entry Module 40811P20 5 Rubber Feet 40811P19 6 Knob 40811P2 7 Valve Body 40811P7 8...

Page 10: ...ide Heater 40811P33 5 Controller 40811P34 6 Drawer Slide 40811P12 7 Solid State Relay 41984P18 8 Handle 40811P26 9 2 Pin Connector 40811P51 10 8 Pin Connector X2 40811P50 11 14 Pin Terminal Block 4081...

Page 11: ...lar 40811P35 6 Motor 40811P31 7 Gearhead 40811P31A 8 Power Supply 40811P40 9 Thermistors 120C 40811P28 10 Tank Assembly 40811G3 11 Bottom Heater 4081132 12 Cover Interlock Bracket 40811P46 13 Panel Sa...

Page 12: ...Tank Stirrer ITEM DESCRIPTION PART NUMBER 1 Thermocouple K Type 10980 2 Main Shaft 40811P13 3 Bearing 40811P29 4 Seal 40811P41 5 Stirrer Assembly 40811G2 6 Stirrer Knob 40811P18 4 3 2 5 1 Inside View...

Page 13: ...13 Accessories ITEM DESCRIPTION PART NUMBER 1 Dispensing Extender 40811P58 1...

Page 14: ...Street Albany New York 12207 USA Toll Free 800 833 2343 USA 518 434 3147 Fax 518 434 1288 USA Email info cmpindustries com Export Email export cmpindustries com Website www cmpindustries com Form 4081...

Reviews: