PROFILE A          15-100%        On/Off             199°F (93°C)             133°F (56°C)           30 mins.             F°           0 “Sea Level”          

PROFILE B          15-100%        On/Off             207°F (97°C)             131°F (55°C)         30 mins.             F°           0 “Sea Level”

PROFILE C          15-100%        On/Off             204°F (96°C)             134°F (57°C)         20 mins.             F°           0 “Sea Level”

PROFILE D     

15-100%        On/Off             207°F (97°C)            131°F (55°C)         15 mins.             F°           0 “Sea Level

Power Supply Interruption

In the event of a power interruption 
(example: electrical outage due to weather),
the computer-controller will remember the
heating cycle and will continue with the
same cycle if the power supply comes 
back online within 15 minutes. If the 
colloid material is not solid, start a normal
cook-down cycle by turning the unit on and
pressing the ENTER button. If the power 
supply is off for more than 15 minutes and
the colloid material is completely solid,
select the cold start cook-down cycle by 
following these steps: With the power on,
press the SET button then the “   ” button
then the SET button twice (display will flash
“DONE”), then press ENTER twice. 

VI. Settings

The factory default profile “A” is set-up 
for Nobilium, Ticonium or Niranium colloid
products used at sea level. If the duplicator
will not be used at sea-level, the altitude
setting must be adjusted. Failure to set 
the Altitude could cause the duplicator to 
deliver poor quality colloid.

Select Pre-Programmed Setting

If not using Nobilium, Ticonium or Niranium
colloid products, review the requirements 
of the duplicating material to be used and
match them to one of the pre-programmed
profile settings below. To select a specific
profile, the duplicator must be in “Standby

Mode” (power on but no heating lights illu-
minated). Press the SET button twice. The
display will show the first option on the set-
tings menu which is the present profile.
Press the “^” (UP) button to find the profile
to be changed.  Press the SET button again
to select the profile. Press the ENTER button
to exit the settings menu.

Develop Custom Settings

If none of the pre-programmed settings
match the requirements of the duplicating
material, then develop a custom profile.
To develop custom profile settings, the
duplicator must be in “Standby Mode”
(power on but no heating lights illuminated).
Press the SET button once. Then press the
“^” button to find the attribute to be
changed. When the desired attribute is 
displayed, press the SET button to select it.
Then press the “^” or “  ” button to change
the setting. Press the SET button to confirm
the change. Press the ENTER button to exit
the settings menu. The attribute setting
options are presented below along with their
maximum and minimum values.

To reset the duplicator to factory defaults,
scroll through the menu options until the
display reads “SET DEF.”  Press the SET
button to return all settings to factory default
settings and “DONE” will be displayed. To
abort changes, press the Enter button to 
re-enter the “Stanby Mode.” Other 
attributes (KP, KI, KD) cannot be edited.

6

Stirrer         Ready     Cook-Down Temp.     Hold Temp.     Cook Dwell     F° or C°       Altitude

Max.

100%           On/Off             207°F or (97°C)        140°F or 60°C        60 Minutes        F°/C°        +255' x 100

Min.

15%            On/Off             190°F or (88°C)        125°F or 52°C     

1 Minute            F°/C°

– 14' x 100

(+ 25,500')

(

1,400')

^  

^  

Display

“STR”          “RDY”                 “SCT”                      “SHT”                  “SCD”          “DEG”           “ALT”

*

* Default settings

Summary of Contents for Ticonium Colloid-Saver 3.5

Page 1: ...for precision temperature control SmartCycle stirring cycle activated only when needed minimizing colloid breakdown High volume no drip no clog dispensing nozzle Pre set programs for virtually any col...

Page 2: ...ty If you have any questions regarding CMP s quality line of dental laboratory equipment and supplies please call toll free 800 833 2343 or 518 434 3147 Fax 518 434 1288 between the hours of 8 00 AM a...

Page 3: ...ed as the Colloid Saver 3 5 duplicator draws about 12 amps at 115 volts As with any electrical device take extra care when handling liquids around electrical connections to avoid accidental shock It i...

Page 4: ...ke sure to add any liquid that may be in the bottom of colloid container The liquid contains ingredients which extend the life of the colloid Replace top cover and rotate it counter clockwise to engag...

Page 5: ...at the very bottom of the tank simply allow it to cool then peel it out Preventive Maintenance Cleaning 1 Clean Colloid Saver 3 5 duplicator approximately once a month or when new duplicating material...

Page 6: ...rofile settings below To select a specific profile the duplicator must be in Standby Mode power on but no heating lights illu minated Press the SET button twice The display will show the first option...

Page 7: ...life VII Troubleshooting 1 Colloid Saver 3 5 duplicator will not turn on A Check wall outlet for power B Check if power cord is connected to duplicator C Check if the main power supply is turned to th...

Page 8: ...P37 2 Precision Laser 40811P48 3 LED Light 40811P38 4 Power Entry Module 40811P20 5 Cover Lock 40811P44 6 Poly Cover 40811P1 7 Cover Interlock 40811P45 8 Cover Handle 40811P43 9 Power Cord Assembly 40...

Page 9: ...ART NUMBER 1 USB Port Connector 40811P39 2 1 Amp Motor Circuit Breaker 65383P7 3 15 Amp Main Breaker 65383P6 4 Power Entry Module 40811P20 5 Rubber Feet 40811P19 6 Knob 40811P2 7 Valve Body 40811P7 8...

Page 10: ...ide Heater 40811P33 5 Controller 40811P34 6 Drawer Slide 40811P12 7 Solid State Relay 41984P18 8 Handle 40811P26 9 2 Pin Connector 40811P51 10 8 Pin Connector X2 40811P50 11 14 Pin Terminal Block 4081...

Page 11: ...lar 40811P35 6 Motor 40811P31 7 Gearhead 40811P31A 8 Power Supply 40811P40 9 Thermistors 120C 40811P28 10 Tank Assembly 40811G3 11 Bottom Heater 4081132 12 Cover Interlock Bracket 40811P46 13 Panel Sa...

Page 12: ...Tank Stirrer ITEM DESCRIPTION PART NUMBER 1 Thermocouple K Type 10980 2 Main Shaft 40811P13 3 Bearing 40811P29 4 Seal 40811P41 5 Stirrer Assembly 40811G2 6 Stirrer Knob 40811P18 4 3 2 5 1 Inside View...

Page 13: ...13 Accessories ITEM DESCRIPTION PART NUMBER 1 Dispensing Extender 40811P58 1...

Page 14: ...Street Albany New York 12207 USA Toll Free 800 833 2343 USA 518 434 3147 Fax 518 434 1288 USA Email info cmpindustries com Export Email export cmpindustries com Website www cmpindustries com Form 4081...

Reviews: