____________________________________________________________________________________
____________________________________________________________________________________
Speeder X1 and X2 Configuration and API Guide
Page 8/36
3)
Sensor information at Power-up and after Resetting
9600
(Baud Rate)
EEPROM PARAMS RESTORED
SPEEDER X1
CMS5012242,RS232
Noptel Oy
ParamDate:2012.04.20
Version :5.00.50 BAF3h
SW Date :Dec 14 2012
SW time :12:51:12
Ubat :10.3 V
READY!
4)
Sensor information <esc>V<cr>
An example of answer for command V:
SPEEDER X1
CMS5012242,RS232
Noptel Oy
ParamDate:2012.04.20
Version :5.00.50 BAF3h
SW Date :Dec 14 2012
SW time :12:51:12
Ubat :11.4 V
OK
The order of lines varies according to the firmware version of the Sensor.
5)
Voltage, current and temperature table (table is for diagnostic purposes only)
An example of answer for command V1 is presented below (units are added):
Description
Value Unit (Not in output)
APD
:02212
[0.1V]
RX temp
:03222
[0.01°C]
RX Anaout
:0156
[mV]
Current
:00201
[mA]
VDD
:05068
[mV]
VAN
:02960
[mV]
VTDC
:04845
[mV]
CP temp
:02540
[°C]
UBAT
:10293
[mV]
UDCDC
:0548
[mV]
VRX
:05161
[mV]
VOSC
:05022
[mV]
VTX
:05138
[mV]
VCP
:03348
[mV]
TXHV
:00257
[V]
TX temp
:02311
[0.01°C]
OK