21
Auto Reset (Hold Time ?)
Auto Reset ?
Not Enabled
Press
or
to change setting.
If Enabled, pressing
or
will take the user to ‘
SETUP
Hold Time ?
’.
If Not enabled the user will return to
SETUP
.
Enabled =
Hold
the torque (and angle) values the length of the
Hold Time
setting after the torque has
been removed and then reset the display to
0
.
Not Enabled =
Hold
the torque (and angle) values until the ZERO
button has been
pressed
and then
reset the display to
0
.
Hold Time ?
Hold Time ?
4
Sec
Press
or
to change value.
Press
or
to confirm and go to
SETUP
(saving changes).
Minimum =
1
,
Maximum =
10
. Default =
4
Display
Inverse ?
Not Enabled
Press
or
to change value.
Press
or
to confirm and go to
SETUP
(saving changes).
Enabled =
Black
digits on
White
background, Not Enabled =
White
digits on
Black
background, when
in the
Measure display
.
WARNING:
WHITE
DIGITS ON
BLACK
BACKGROUND
WILL
REDUCE
THE
OPERATIONAL BATTERY LIFE BY APPROXIMATELY
65%
.
Torque CRS
Distance ?
31.8
mm
Press
or
to change value.
Press
or
to confirm and go to
SETUP
(saving changes).
Minimum =
0.0
, Maximum =
999.0
. Default: - NorTronic
®
50 & 200 =
31.8 mm
, NorTronic
®
330 =
35.0 mm
.
Summary of Contents for NorTronic 43500