Clock settings and RTC alarm reset RESET
Active alarm
Controller will show active alarm alternated with the temperature inside the unit.
How to export data and alarms to USB
1. Insert the
USB
key
2. Push the
DATA
key for more that 3 sec.
3. Controller starts uploading data to
USB
4. At the end the following message will be displayed:
a.“
End
” if everything it’s ok
b.“
Err
” if exporting has not taken place.
Alarms
The controller memorizes the last 100 alarms happened, together with their start and finish
time. It’s possible to export the alarms as described in the previous chapter.
1. Push the UP key once, to acces the menu.
2. The display shows H∩, then push SET.
3. The parameters for setting time and date occurs. To set the parameter push SET, push the
UP or DOWN button to change the parameter. Confirm by pushing SET.
The time and date parameters:
Hur: hour
∩in: minutes
Udy: weekday
dAy: date
∩on: month
yEA: year
To exit:
Press
SET
+ UP
keys for about 10 sec. or wait 60 sec.
Alarm signals
Message
Cause
“PF1”
Regulating probe TR3 failure
“HA1”
TR3 High Alarm
“LA1”
TR3 Low Alarm
“dA”
Door Open Alarm
“CA”
Serious Alarm
The alarm message is displayed until the alarm condition is recovered.
All the alarm messages are showed alternating with the temperature in the unit.
Except for the “PF1” which is flashing.