13
Norsonic Nor1218
Instruction Manual
Connecting the Nor1218 to
Sound Level Meter Nor131 or
Environmental Noise Meter Nor139
The Nor1218 is designed to be used with the Sound
Level Meter Nor131 or Nor139. These instruments
support all the needed corrections for the two types of
microphones. This chapter discuss how you set up and
use the Sound Level Meter.
In the SETUP – 1 (Instrument menu) – 4 (Input menu),
you select which type transducer you have connected.
Select either “1218 Vertical” for Nor1218 for vertical fre-
quency correction to be applied and “1218 Horizontal”
for horizontal frequency correction.
If the large windscreen Nor4576 is fitted above the
standard windscreen supplied with the Nor1218 an
extra frequency compensation is needed. This is only
possible in the horizontal mode. Select 2: corrections
in the input menu and then 1: large windscreen in the
next submenu to turn this correction on. Please note
that this correction is always set to OFF each time the
instrument is turned off.
It is not possible to use large windscreen if vertical mode
is selected. Adding the extra windscreen is required in
some applications when measuring on wind turbines.