Record setup menu
Record trigger setup
Setup for a recording
Press
SETUP
>
8
and menu for specifying the record-
ing will be shown.
Select the duration for each recording in seconds. Enter
a value in the range 0 to 9999 seconds by keying in the
value or by using
INC
or
DEC
to modify the displayed
value. The value 0 is used for selecting a recording which
lasts to the end of the measurement.
Select the required format for the recording by se-
lecting the number of bits and appropriate sampling
frequency. All recordings are made in a standardised
WAV-format which allows most media-players to play the
recorded file.
Setup how to start a recording
A press on
STORE/RECORD
will start a recording with
the selected duration.
If the trigger option is installed (Option 16), trigger-
ing from an external signal or from a noise event is also
viable. Press
SETUP
>
6
for triggering and then select
2
for record triggering. You have to select between the
following alternatives:
Manual trigger
Set the external and level triggering to “OFF”. Select
the time delay between the trigger command and the
execution. A value in the range -99 to 99 may be used.
-99 means that the record starts 99 seconds before
RECORD
was pressed. Note that a measurement has
to be running before you are allowed to store a record!
Key in the value or use
INC
or
DEC
to modify the dis-
played value.
External trigger
Set External to “ON” by the use of the
INC
or
DEC
keys
if you want to enable triggering by an external signal
to the digital input terminal. After pressing the
START
key, the instrument will start waiting for an externally
supplied binary voltage signal (3,3 volt CMOS). The
signal has to be applied to DI-1, pin no 8 on the general
I/O-socket. See the specification section for further in-
formation.
The remote hand switch Nor263A is designed for
external trigger of either remote start of the meas-
urement process or remote start of an audio record-
ing. The trigger conditions is set in the trigger menu;
SETUP
>
6
. After selecting the trigger condition the
digital IO must be configured to use the remote hand
switch Nor263A. You may skip the configuration, since
the push button will work regardless of the setting.
However, to achieve proper operation of feedback LED
mounted on the handswitch, you need to set this in the
digital IO menu. Push
SETUP
>
1
(Instrument menu) >
2
(IO / Print) >
1
(Digital I/O). The LED is connected to
DO-1. Please note that the QC (Quality Control) func-
tion in the Reference Curve menu
SETUP
>
4
(Freq
mode) >
1
(Reference curve) must be turned off, un-
less is DO-1 and DO-2 reserved for the QC function.