93
Norsonic Nor140
Instruction Manual
Insert a reference tone as a recording
When listening to a recording, it may be required to
make the playback with the same actual sound level at
the listener’s ears as the original sound was at the spot
of the actual measurement. In such cases a reference
tone with a pre-defined level may be recorded dur-
ing the measurement, and later replayed through the
listener’s loudspeaker system at the spot of the replay.
The reference tone is activated by use of the
CAL
key during a running measurement provided that it
has been enabled in the Ref.tone menu. The menu is
displayed by pushing
SETUP
>
1
(Instr.) >
8
(Ref.tone).
The reference tone feature is enabled by using the
cursor keys to move the cursor to the upper position
and selecting “ON” using the
INC
or
DEC
keys.
The preset type, gain and excitation time of the
reference tone are edited in the same menu. The type
is either PINK noise or a SINE wave. If the SINE is
selected, the actual frequency is set at the lower cur-
sor position. The default level of the reference tone
is 120 dB. This may be lowered by choosing a gain
between 0 and -50 dB; hence producing a reference
tone in the 70 – 120 dB range. The excitation time, or
duration, of the reference tone after activation by the
CAL
key is also selected in this menu. The duration is
variable within the 1 – 60 second range.