103
Norsonic Nor140
Instruction Manual
from 200 V to about 70 V. The microphone sensitivity
will then be reduced by 10 dB and the instrument will
be able to measure peak signals up to 150 dB.
The change in the polarisation voltage will lead to
a small change in the frequency response for the mi-
crophone. This change is automatically compensated
when the extended measurement range is selected.
Since this correction is adapted to the microphone
Nor1225, the extended measurement feature should
not be used with other types of microphones.
To activate or de-activate the extended measure-
ment range, press
SETUP
>
1 > 4 > 1
. This submenu is
only possible to access if STANDARD input is selected
in
SETUP 1 > 4
. An “H” is displayed in the lower part of
the display when extended range is selected.
Normally a recalibration of the instrument is not
needed. However, it is good practice to check the sen-
sitivity of the instrument. This can easily be done by
applying the usual sound calibrator. The sound level
meter shall indicate the usual level as stated for the
calibrator also when the extended measurement range
is selected.
Marks on the lower line of the display
Marks for the applied corrections are
found on the lower line of the display:
dB
The signal strength is indicated as a level
in decibel. The reference level is normally
20 µPa for sound pressure levels.
EU
Engineering unit: The signal strength is
indicated in a generic linear unit. The
actual unit could be voltage referring to the
voltage on the input terminal or ms-2 if an
accelerometer is connected to the input.
#
Numerical keyboard. The number printed
on the keys are entered if you press one of
the keys on the keyboard
E
A numeric value has been entered. The in-
strument expects that you press
ENTER
to
confirm the number.
N/H
Indicates
N
ormal or
H
igh measurement
range. (Polarization voltage dependant.)
?
A key is pressed that the instrument does not
understand.
Additionally there are marks indicating the applied
corrections found on the lower line of the display:
R
Random incidence correction ON
W
Wind screen correction ON
G
Preamplifier correction ON
S
Self-noise correction ON
E
Polarization voltage off - Electret micro-
phone
Summary of Contents for nor140
Page 4: ......
Page 16: ......
Page 17: ...nor140 SOUND ANALYSER ...
Page 18: ......
Page 212: ...194 ...
Page 218: ......