152
Chapter 24
Building Acoustics
Note
that you have to be in the menu for
entering values before you are able to select
another mode.
V volume of the receiving room [m
3
]
A
0
reference absorption area equal to 10 m
2
.
Sound excitation for facade measurement
−
s
The relations between the quantities are given by the
following equations:
The values are calculated for each octave band. For
D
nT
and D
n
, a single-number value is also calculated.
The value corresponds to the 500
H
Z
value for the ref-
erence curve after shifted it in accordance with the
method specified in ISO 717-1 for octave-band levels.
The reference curve is shifted in one-dB steps until the
unfavourable deviations between the octave-band re-
sults and the reference curve is as large as possible,
but not more than 10 dB for octave bands or 32 dB for
one-third octave bands.
Noise excitation
The applied method requires broadband noise to be
used for the excitation. The noise may be generated
by the instrument if option 10, Noise generator, is in-
stalled. Sometimes it will be convenient not to have a
cable between the instrument and the loudspeaker for
the excitation. This may be achieved by using a radio-
transmitter for the noise signal (contact Norsonic for
more information) or to use a power amplifier with an
internal noise generator such as the Nor280. Another
alternative is to play a music-CD with recorded noise.
Norsonic can supply a CD with the required excitation
signal (Nor1033)
The menu for airborne sound insulation is also
used for facade measurements