156
Chapter 24
Building Acoustics
You are prompted to accept the recalled or meas-
ured values as a part of the averaged result. Press
ENTER
(alternatively
STORE)
to accept or
EXIT
(al-
ternatively
DEL)
if you don’t want to accept. You may
repeat the loading of data to obtain the averaged re-
sults from more measurements. The number of meas-
urements that has been accepted for the averaging is
displayed by a number N. Press the
TBL
key to display
the averaged levels.
Reverberant position Sound Pressure Level
The next set of values to be entered is the sound pres-
sure levels in the reverberant field of the room Lr for
the octave bands in the frequency range 31,5 Hz to 8
kHz. The distance to the corner measurement position
and to any source inside the room should be minimum
1,5 metre, and the height above the floor should be
between 0,5 and 1,5 metre. Two measurement posi-
tions should be measured in the reverberant field with
a distance of minimum 1,5 metre from each other. Re-
call a measurement by using the
RECALL
key or press
START
for starting a measurement. If measured, the
values are automatically stored.
You are prompted to accept the recalled or meas-
ured values as a part of the averaged result. Press
ENTER
(alternatively
STORE)
to accept or
EXIT
(al-
ternatively
DEL
) if you don’t want to accept. You may
repeat the loading of data to obtain the averaged re-
sults from more measurements. The number of meas-
urements that has been accepted for the averaging is
displayed by a number N. Press the
TBL
key to display
the averaged levels.
Reverberation time
The final set of values to be entered is the reverbera-
tion times. The acoustic losses of the receiving room
used in the calculation may be obtained by measur-
ing the reverberation time or by recalling earlier meas-
ured values. See the earlier paragraph under airborne
sound insulation describing measurement of reverber-
ation time. If the optional noise generator is installed
it may be used for the excitation, otherwise impulse
method has to be used. To start a measurement, press
START
when the field cursor is in the field for rever-
beration time marked “T”. After the measurement you
may inspect the resulting reverberation table by press-
ing the
TBL
key.
Please note that the reverberation times are meas-
ured in the octave bands from 63 Hz to 8 Khz. The 31,5
Hz band is not considered according to the ISO 16032
Standard.
Display of results
When the required information has been entered, the
cursor will automatically be placed in the field for cal-
culation of the final result. If you are satisfied with the
entered values press
ENTER
, or if you want to make
any correction, move the field cursor to the required
field and enter the corrected values or press
DEL
to
clear averaged values.
When you move the field cursor to the field “Calc
results” and press
ENTER
, the following functions are
calculated for both the A- and C-weighted functions:
• The service equipment normalised sound pres-
sure levels L
eq, nT
, L
Smax,nT
and L
Fmax,nT
•
The service equipment standardised sound pres-
sure levels L
eq, n
, L
Smax,n
and L
Fmax,n