166
Chapter 25
Speech Intelligibility by STIPA
Performing a measurement
Set-up
Set the sound source (artificial mouth or suitable test
loudspeaker) in the normal position for the speaker. If
a sound system is being used, place the sound source
on the axis of the appropriate microphone at the normal
speaking distance (measured from the lip-circle for the
artificial mouth or acoustic centre of the loudspeaker)
and direct it in the normal speaking direction. Connect
an appropriate audio-player to the sound source.
Set the speech level
Switch on the sound level meter and ensure that it is
properly calibrated. Play track 2 with the STIPA excita-
tion signal and adjust the level to the required speech
level, normally 60 dB at 1 m from the sound source.
(66 dB at 0,5 m) Alternatively, track 1 with the sine ex-
citation may be used – both should produce about the
same A-weighted level.
After setting the speech level, play the STIPA exci-
tation signal on Track 2 during the measurement. (70
minutes).
The standard IEC 60268-16 recommends that the
excitation spectrum is correct within ± 1 dB for the ap-
plicable frequency range.
Select the STI-mode
The STI mode is one of the modes of operation for
the sound level meter. By selecting this mode, the
instrument will automatically be configured for STI-
measurements. For selection, press
MODE
after
switching the instrument on, and select STI by
pres-sing the numeric button
5
. After the selection, the
instrument is ready for a STI-measurement. A “$”-sign
in the upper left corner of the display indicates the STI-
mode of operation.
The instrument screen will show the level versus
frequency display. Time constant will be F. The cursor
will be positioned for indication of the A-weighted level.
Making a measurement
Start a measurement by pressing the
START
button. The measurement lasts for 13 seconds. During
the measurement the level versus frequency display
will be shown together with an indicator of an ongoing
(running) measurement. After the measurement is
ended, the main STI display will be shown after a short
calculation period.
For the first measurement after the STI-mode is
selected, the Noise correction will be selected off and
the cursor will be placed above the “Off” field.
The measured result will be indicated as a STI-
value together with an assessment of the intelligibility.
The STI-value and the assessment are according to
the STIPA-procedure in IEC 60268-16 for male voice.
The corresponding CIS-value (Common Intelligibility
Main result display
Display during
measurement of STIPA