Norsonic Nor140
Instruction Manual
FFT
measurement mode
Introduction
When the FFT option 14 is installed, the instrument
may make a narrow-band frequency analysis of the in-
put signal and calculate the auto spectrum. The signal
is sampled with a sampling frequency of 24 kHz and
a
F
ast
F
ourier
T
ransformation based on 2
14
= 16384
samples is performed. This allows the signal to be ana-
lysed with a constant frequency resolution of 1.46 Hz.
The signal strength may be displayed in logarithmic
[dB] or linear units [engineering units EU].
How to select the FFT-mode
To put the instrument in FFT-mode of operation, press
the key
MODE
and select
6
for FFT. If you want to return
to the normal mode of operation, you have to press the
key
MODE
again and select
1
for Normal.
Making a measurement
A measurement in FFT mode may be started as a nor-
mal level measurement by pressing the
START
button.
If Option 16 – extended trigger is installed, the same
trigger possibilities apply as in normal mode of opera-
The FFT-mode is selected in the mode menu.
tion: the measurement may be started at a certain time
of the day or if the level goes above a certain level.
The level triggering is based on the normal weighting
network or fractional-octave band level.
The measurement duration may be set as time or
as a number of single measurements used to obtain
an averaged auto-spectrum. See figure overleaf. NA
indicates the number of single measurements in the
averaged value. The averaged result is continuously
updated on the screen during the measurement. The
measurement may be stopped before the selected
number of periods is reached by pressing
STOP
. The
duration of each single FFT-period is 0,685 s.