11
Norsonic Nor140
Instruction Manual
11
Low Voltage Situations
When the combined battery voltage drops below 3.9 V,
a battery low indicator appears in the display and the
instrument will start to shut itself off. Any ongoing
measurement will be terminated and stored in a
directory called
BATLOW
. Memory contents are
retained without the use of electrical power (flash
memory). Upon installing fresh batteries, the instrument
will at start-up ask the operator for the confirmation
to store the previous measurement in the normal
measurement directory.
If the instrument is connected to an external
DC-source, the external-source-voltage will be
displayed (EV). Should the external power fail during
a measurement, without any internal batteries installed
(or the installed batteries have no power left), the
instrument will be turned off immediately without
No recharging.
Connecting an external DC-
source (11–15V) to the instrument, will not
charge rechargeable batteries, but power
the instrument in lieu of the internal batteries.
Polarisation voltage
The polarisation voltage setting
menu opens up for the use of pre-
polarised microphone cartridges.
SETUP
>
1
>
4
storing the ongoing measurement. However, as the
instrument automatically make a backup storage every
2 minutes, the last file stored will include the correct
results except for maximum the 2 last minutes prior to
the power failure. Upon return of the external power,
the instrument will automatically start to measure as if
the
START
key had been pressed.
If powered from internal batteries and left un-
attended and unoperated, the Nor140 will switch itself
off after ten minutes. However, this does not apply if
the instrument is measuring (including being paused
during a measurement) nor when powered from an
external source. The automatic switch off feature
may, however, be disabled if required for long-term
monitoring purposes. Press
SETUP
>
1
(Instr.) >
9
(Misc.par.) >
9
(PowDown), and use the
INC
and
DEC
to alter the power off field from
ENABLED
to
DISA-
BLED.
This feature may be turned off in
SETUP 1 > 9 > 9
and alter the
START
field from
ENABLED
to
DISA-
BLED
. This field is by default set to
ENABLED
. It may
be disabled if the user wants to monitor it and when the
power to the instrument is turned off / tampered.
Time elapsed
since battery
replacement
Combined
battery voltage
as an f(t)
Battery voltage
Battery voltage is presented graphically as f(t). Press
the
BATT
key to produce this display and press again
to exit the display.
The external supply
voltage is shown
when connected,
else battery voltage