NorthEast Monitoring, Inc. NEMM019-Rev-O
Page 18 of 35
DR200/HE & DR300 Operator’s Manual
Event Recording: Event Recording: Patient Operating Instructions
Step 5:
At this time the
technician may ask you
to erase your events by
pressing either of the
green arrow buttons
down for 3 seconds,
but you SHOULD
ONLY ERASE EVENTS WHEN
INSTRUCTED TO DO SO. Once events are
erased, you will see that the first number is
changed to 0 as there are no longer any events
stored on the recorder. You are now ready to
record new events.
Wireless Event Transmission:
If your DR300 digital recorder is set up to do
automatic wireless transmissions, you do not
need to send events manually. Instead, when
the recorder comes in contact with the North-
East Gateway box, that has been provided to
you, it will automatically send any events that
are saved every 20 minutes.
Cleaning
You can clean the outside of the recorder with a
soft cloth.
LCD Screen Messaging
Error messages may occur during recording. If
you see an error message that is not listed on
this page call your receiving center for further
instructions.
•
If the time and event tally appears on the
screen,
this is normal and means that the
recorder is functioning appropriately.
•
If the screen is blank,
this means that the
recorder is not working. It is possible that
the battery no longer has enough power. Try
putting in a fresh battery, but if this does not
solve the problem, call your receiving cen-
ter for further instructions.
•
Battery LOW:
Battery power is running
low. When this message first appears, you
will have about 5 days of recording time
left on your battery. In order to ensure that
your next transmission is successful, b
e
sure to replace the battery before you send
any transmissions to the receiving center.
•
Battery FAILURE:
Recording has
stopped. Replace the battery as soon as pos-
sible.
•
LEAD LOOSE:
This error will occur when
there is a problem with the patient hook-up.
The problem may be with an electrode, a
lead, or the cable that connects the leads to
the recorder. Please check all of the connec-
tors to be sure that they are still in place. If
you continue to see this error after 10 sec-
onds, call the receiving center for further
assistance.
Batteries
Your battery should last for the duration of
your test, but if you need to change the battery,
you can do so at any time and not lose any
event data stored on the recorder.
To replace the battery, insert a blunt object (for
example, pen, coin or non-pointy tool) in the
space between the battery and the top edge of
the recorder. Press gently to easily remove the
battery.
To insert a fresh battery into the battery com-
partment, be sure to orient it as indicated in the
diagram inside the compartment. The battery
sits loosely in the compartment.
Note: Do not leave batteries in the recorder
for extended periods (more than two weeks)
when the recorder is not in use.
89
12:45
Recordings
0 / 100