NorthEast Monitoring, Inc. NEMM019-Rev-O
Page 25 of 35
DR200/HE & DR300 Operator’s Manual
DR300 Tablet App: Live ECG Viewing
•
Save Recorder Settings on Tablet - Saves the
settings you just entered onto the Tablet for
future use
•
Update Time - Will automatically update the
time and date on the recorder to match the tablet
The Wireless Settings
The Wireless Settings menu is where you can turn
Wireless On or Off. When Wireless is turned On,
the DR300 recorder will transmit saved event or
Holter through a paired NorthEast Monitoring
Gateway box.
The other items on the Wireless Settings show you
whether the recorder is paired and how your
recorder is set up for transmitting.
Note: Refer to the DR300 Gateway and Socket
Technical Manual, NEMM046, for information
on how to set up and run the Wireless feature.
Locked Recorder
Locked recorders cannot be updated or started with
the Tablet. To update settings on a locked recorder,
you will need to first unlock the recorder by going
to Settings > More > Menu Lock, and entering:
•
217 to lock the menu
•
151 to unlock the menu.
Live ECG Viewing
If the recorder is in Event or Holter recording mode
(not Both), at any point during recording you can
view the live ECG signal on the recorder and/or a
Tablet with the DR300 App turned on.
In order to view the live ECG signal on the
recorder, start the DR300 Tablet App, and then
press the arrows in the following order - up, down,
up, down. At that point the live ECG signal should
appear on the recorder screen and will transmit to a
tablet when available.
To stop viewing, press the record arrow in the
upper-left corner of the Tablet App. On the
recorder, press the down-arrow to stop live viewing.
Troubleshooting
If DR300 Android App is not connecting with the
DR300 recorder that you want to configure and/or
start, you may want to try one or more of the fol-
lowing:
•
Close and restart the DR300 Tablet App
•
Make sure you have the correct recorder Serial
No.This can be found on the DR300 About
screen and in the battery compartment on the
back of the DR300.
•
Make sure that only one DR300 recorder is open
to its Splash screen.
•
Check to make sure that the DR300 is Paired
with the tablet. You should see the recorder
Serial No. on the Tablet’s Bluetooth screen.
•
If open, close the DR300 App and restart it.
•
If Paired, then turn off the Bluetooth and then
turn it back on again on the tablet. Also close
and restart the DR300 App. Once restarted,
make sure that the recorder’s Serial No. is still
visible on the Paired list.
•
Restart the DR300 recorder be removing and
reinserting the battery.
•
Make sure that there are no unerased events on
the recorder.
Additional Notes:
•
If you switch the recorder from Event to Holter
recording mode, the recorder will restart and
you may need to enter the Patient ID two times.
•
If you turn of the tablet’s Bluetooth radio will
the DR300 Tablet software is not closed, the
software will go into a continuous mode looking
for the Bluetooth. Close the software and restart
to fix.
•
The DR300 recorder will restart, if you close the
software while connected to the recorder.