background image

TCM Endo III 

 

NOUVAG AG 

 

 

No. 31817   29/05 

1  Description 

1.1  Operation and functions 

The TCM Endo is a microprocessor controlled endodontic electronic motor system. 
The slow speed, high torque rotary instrument uses an electronic torque controlled motor to achieve a 
faster  and  easier  root  canal  preparation.  The  speed  is  held  constant  under  varying  loads  until  the 
selected torque level is reached. 
The protection mode (AP) provides an automatic protection against file breakage. When the file reaches 
the selected torque level, the motor reverses immediately to protect the file, and then returns to forward 
to finish the root canal preparation. With use of the (AP) mode, file stress is greatly reduced. 

1.2  Technical Data 

Voltage: ............................................................ 115V

 / 50-60Hz or 230V

/ 50-60Hz (see device label !) 

Fuses: ..................................................................................2 x T2A (for 115V

) and 2 x T1A (for 230V

Power: ..............................................................................................................................................40 VA 
Type:  ......................................................................................................................Application part type B 
Speed:  ....................................................................................................................... 1,200 – 16,000 rpm 
Dimension: ..................................................................................................................90 x 100 x 170 mm 
Net weight:  ..........................................................................................................................................2 kg 

1.3  Operating Environment 

Relative Humidity: ....................... max. 80% 
Temperature: ..............................10 to 40°C 
Barometric Pressure:......... 800 to 1060hPa 

1.4  Transport and Storage Environment 

Relative Humidity:............................. max. 90% 
Temperature:..................................... 0 to 60°C 
Barometric Pressure: ..............700 to 1060hPa 

1.5  Description on device 

 

: Attention: Read instructions first 

 

: Application part Type B 

1275  : Sign of EU-Conformity 

 

 

: Certified by the Canadian 

Standards Association (CSA) for 
Canada and USA

 

 

: Old electrical and electronic 

equipment must be disposed 
separately and may not be 
included in regular domestic 
waste. 

Summary of Contents for TCM Endo III

Page 1: ...NOUVAG Made in Switzerland Nr 31817 www nouvag com Operation manual GB TCM Endo III ...

Page 2: ...d Storage Environment 2 1 5 Description on device 2 2 Safety measures 3 3 Shipping contents 3 4 Set up 3 5 Operation 4 5 1 The TCM Endo III unit 4 5 2 The footswitch 5 5 3 Automatic Limiter AL 5 5 4 Automatic Protection AP 5 5 5 Memory 6 6 Disinfection Cleaning and sterilization 6 6 1 Control unit and footswitch 6 6 2 Mikromotor Endo 7 7 Troubleshooting 7 8 Disposal 7 ...

Page 3: ...l preparation With use of the AP mode file stress is greatly reduced 1 2 Technical Data Voltage 115V 50 60Hz or 230V 50 60Hz see device label Fuses 2 x T2A for 115V and 2 x T1A for 230V Power 40 VA Type Application part type B Speed 1 200 16 000 rpm Dimension 90 x 100 x 170 mm Net weight 2 kg 1 3 Operating Environment Relative Humidity max 80 Temperature 10 to 40 C Barometric Pressure 800 to 1060h...

Page 4: ... nor can Nouvag AG be held liable for any injuries to persons or animals in any case when the device is miss used or not operated applied or maintained in strict accordance with the user owner instructions set out in the operating manual In the event of any doubt or question the user is to contact Nouvag AG or its lawful representative for clarification or assistance 3 Shipping contents 1520 Contr...

Page 5: ...r reduce torque Rev Forw To change motor direction of rotation a warning tone will sound when the motor operates in reverse mode 1 1 Press this key if using handpiece 1 1 Speed from 1 200 to 16 000 rpm Torque 1 2 3 4 5 7 10 13 16 19 Nmm 8 1 Press this key if using contra angle 8 1 Speed from 150 to 2 000 rpm Torque 2 4 6 10 15 20 25 30 40 50 Nmm 16 1 Press this key if using contra angle 16 1 Speed...

Page 6: ...20 2 Pressing the keys and will increase and decrease the torque level respectively 3 By pressing the ATC SET briefly once more the digital display switches back automatically 5 4 Automatic Protection AP The automatic protection AP mode is a special function against file breakage When the selected torque level is reached the motor reverses immediately to free the file and then returns automaticall...

Page 7: ...olving agents for cleaning Autoclave material in transparent packaging Do not fill the sterilization bag more than 80 Autoclave material at 134 C maximum Sterilized material should be stored and tagged with sterilization date 6 1 Control unit and footswitch The control unit and the pedal do not come into contact with patients Only clean the surface of the unit do not use harsh cleaners or solvents...

Page 8: ...properly Insert plug and check fitting Wrong operating voltage Check mains voltage output please see operating voltage on the rating plate Footswitch fails to function Pedal not connected Plug pedal cable into rear of control unit Control unit not turned on Turn main switch ON OFF Micromotor does not run Motor not turned on Turn motor with Motor key on or with footswitch Motor not connected Connec...