Interfering Substances
Glucose Interferences:
The Nova StatStrip Xpress-i Glucose and ß-Ketone Hospital
Meter exhibits
no
interference for glucose from the following
substances up to the following concentration levels:
Substances
Conc
mg/dL(mmol/L)
Acetaminophen
10.0 (0.66)
Ascorbic Acid
10.0 (0.57)
Bilirubin
15.0 (0.26)
Cholesterol
500.0 (12.9)
Creatinine
6.0
(0.53)
Dopamine
10.0 (0.53)
Ephedrine
0.9
(0.055)
D(+) Galactose
350.0 (19.4)
Hematocrit (RBC) 65%
Ibuprofen
48.0 (2.33)
L-Dopa
100.0 (5.07)
Introduction
Substances
Conc
mg/dL(mmol/L)
D(+) Maltose
240.0 (6.66)
D(+) Maltotetraose 240.0 (3.6)
D(+) Maltotetriose 240.0 (4.76)
Methyl-Dopa
1.0
(0.042)
Oxygen All Conc
Salicylate
30.0 (1.87)
Tetracycline
30.0 (0.62)
Tolazamide
15.0 (0.48)
Tolbutamide
45.0 (1.67)
Triglycerides
750.0 (8.78)
Uric Acid
20.0 (1.05)