1
Important Information for Control Solution
• Use only the Nova Max Control Solution for the test.
• Check the expiration date on the control solution vial. Do not
use control solution past the expiration date or you may get
inaccurate results.
• Store only for 3 months after first opening. When you open
a new vial of control solution, count forward 3 months and
write that date on the label of the control solution vial. Dis-
card any remaining solution after the date you have written
on the vial.
• Store the control solution tightly closed at room temperature
below 86°F (30°C). Do not refrigerate or freeze.
• Shake the control solution well before using.
Caution:
The Nova Max Control Solution range printed on the test
strip vial is for control solution only. It is used to test the perfor-
mance of the monitor and test strip. It is not a recommended range
for your blood glucose level.
If your control solution test results continue to fall outside the
range printed on the test strip vial:
• The Nova Max Link Blood Glucose Monitor may not be work-
ing properly.
• Do not use the monitor to test your blood.
• Call Customer Service at 1-800-681-7390.
Running Control Solution