3 Installation
3.4 Set up for Initial Use
MOVIA, 2-Axis Scan Head – User Manual, 1040–0004 Rev. 03
18
controllers. We can also create correction tables for custom configurations on
request.
In the case of the ScanMaster Controller, you would use Cambridge Technology's
CalWizard software and follow the procedure outlined in the CalWizard manual
and the Calibration Applications note.
After calibration has been completed, installation of the scan head is complete.
3.4
SET UP FOR INITIAL USE
Delay parameter settings are variable because each application’s requirements
and the controllers that are used fluctuate. Use the delay settings outlined
in Table 3 as a starting point, then fine tune the settings as needed.
For additional information on delay settings, see Cambridge Technology’s
CalWizard manual P0900-0157, Appendix A.
Table 3: Recommended Delay Settings for Initial Use
MOVIA 10mm Scan Head
Mark Delay
130
µ
s
Jump Delay
260
µ
s
Poly Delay
65
µ
s
Laser On
Laser-dependent
Laser Off
Laser-dependent