Nsense probe dimmensions 

 

 

2.Configuration  

                                   

Already completed by Novatron prior to delivery  

 
With the next 4 steps you assign the input sensors to output channels 

Step 1:

 Connect nSens to Quantadat and start detection 

 

Step 2:

 Assign nSens to measuring points A to D 

 

Step 3:

 Assign measuring points and parameters to channels 1 to 4 

           Channel 1 equals analog output 1 and display line 1

 

 

Step 4:

 Configure analog signals according your specification  

             

analog output CH1 linear scale  4-20mA, -20 to +80 °C 

             analog output CH 2 linear scale 4-20mA, 0 to 100% rh   ( supplied already completed) 

Step 5: 

If the parameters need to be changed from  standard configuration in Step 4 above 

then please follow these instructions from the three front menu keys 

UP, DOWN, ENTER:-

 

 

              
 

Set up of g/ Kg ( mixing ratio)                     Scaling range of output in g/Kg (mixing ratio) 

 ENTER                                                          ENTER 
‘Measurement’                                                DOWN   
ENTER                                                            

‘Analogue Outs’ 

DOWN                                                             ENTER 
‘Channel 1’                                                      DOWN 
ENTER                                                            

‘Analogue Out 1’ 

‘Measurement Point A’                                    ENTER 
DOWN                                                             DOWN 

 

‘Mixing Ratio’ 

                          

‘Range L 0.00 g/ Kg’ 

DOWN                                                             DOWN 
EXIT                                                                

‘Range H 100.00 g/Kg ( change as required) 

                                                                         EXIT 

            

Note: If you get stuck in the menu just down-power momentarily to return to 
standby mode. On first power up always press ENTER. 

The  following steps 1 

– 4  are only needed if you need to configure the 

Quantadat differently from the original specified settings shown in the previous 
section 2 ( direct replacement for old TR200 Triplex system) or if the 
configuration, carried out by Novatron prior to delivery,  has been altered in 
error . Otherwise, please ignore the following steps, you are now ready to skip 
to  section 3 Calibration.

 

 

 

 

 

Summary of Contents for Novasina Quantadat TRIPLEX

Page 1: ...Novasina Quantadat TRIPLEX humidity system with nSens probes Set up commissioning manual...

Page 2: ...gs can be interchanged without any issue Configuration may be required depending on the application eg the range of measurement the output parameter or the mA analogue output However the standard conf...

Page 3: ...adat Transmitter dimmensions Nsense probe Green Brown White Power 24 V Analogue outputs 4 20mA CH2 20 to 80 C CH1 0 to 100 rh Power supply 24 V ac dc Output CH2 4 20mA Temp 20 to 80 C Output CH1 4 20m...

Page 4: ...se instructions from the three front menu keys UP DOWN ENTER Set up of g Kg mixing ratio Scaling range of output in g Kg mixing ratio ENTER ENTER Measurement DOWN ENTER Analogue Outs DOWN ENTER Channe...

Page 5: ...been recognized already this message appear You can choose Yes to continue and discard previous sensors Quantadat is now searching for connected sensors Step 2 Assign nSens to measuring points A to D...

Page 6: ...parameters to channels 1 to 4 Analog Output equals display line 1 etc Enter Choose Source Measuring point A to D Choose measuring point as assigned in step 2 des Messpunktes A to D gem ss Schritt 2 If...

Page 7: ...put range On Fail Output value setting in mA or V in case of a sensor failure e g interrupted communication with probe Adj Clear Deletion of adjustment This menu command is only shown if the analogue...

Page 8: ...is 2 relative humdity and therfore the maximum difference you should see between all three probes should be 4 relative humdity Using the SC Sensor Check calibration salts Novasina calibration service...

Page 9: ...pletion of installation Installation and comissioning is now complete and you are ready to put the Triplex Quantadats into operation For any further assistance please contact Novatron Scientific Ltd t...